Required temperature and humidity for the oxygen generator
Operating temperature: Between approximately 0
℃
and 40
℃
- If the product is operated under surrounding temperatures with significantly large fluctuation
by external factors, dew condensation occurs thereby shortening the filter life and causing a
product failure.
- If the product is operated at 40
℃
or higher temperature, oxygen generation capability of the
product can be degraded.
Operating humidity: Less than 90%
- If the product is operated at 90% or higher humidity, oxygen generation capability of the prod-
uct can be degraded.