Overload Protection: 0.2A / 250V and 20A / 250V Fuse.

Maximum Input: 200mA DC or 200mA AC rms on mA ranges, 20A DC or AC rms on 20A range.

Overload Protection: 0.2A / 250V and 20A / 250V Fuse. Frequency Range: 50 to 400 Hz

Maximum Input: 200mA DC or 200mA AC rms on mA ranges, 20A DC or AC rms on 20A range.

Resistance

Input Protection: 250V DC or 250V AC rms.

Capacitance

Input Protection: 250V DC or 250V AC rms

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V DC or AC rms. 

  5 

  6 

4.

V,

Ω

,CAP, Hz, TEMP. input jack

5.

mA input jack for mA DC or AC measurements

6.

20A (positive) input jack for 20A DC or

AC measurements.

7.

Data Hold pushbutton

8.

/

  pushbutton.

9.

Backlight pushbutton.

10. Power pushbutton: the power button turns the meter

ON or OFF.

SPECIFICATIONS

The instrument complies with: 

EN61010-1.

Insulation: 

Class2, Double insulation.

Overvoltage category: 

CATII - 1000V.

Display:

2000 counts LCD display with function indication.

Polarity:

Automatic, (-) negative polarity indication.

Overrange:

“OL” mark indication.

Low battery indication:

The  “BAT”  is  displayed  when  the 

battery voltage drops below the operating level.

Measurement rate: 

2 times per second, nominal.

Operating environment:

0 oC to 50 oC (32 oF to 122 oF) at

< 70 % relative humidity.

Storage temperature:

 -20 oC to 60 oC (-4 oF to 140 oF) at <

80 % relative humidity.

For inside use, max height: 

2000m

Pollution degree:

 2

Power:

One 9V battery , NEDA 1604, IEC 6F22. 

Dimensions:

195 (H) x92 (W) x 38 (D) mm

Weight: Approx.:

380g.

Accuracy is given at 18 oC to 28 oC (65 oF to 83 oF), less
than 70 % RH

DC Voltage

Range

Resolution

Accuracy

200.0mV

0.1mV

+0.5% of rdg + 2 dgts

2.000V

1mV

20.00V

10mV

200.0V

100mV

1000V

1V

+0.8% of rdg + 2 dgts

Input Impedance: 10M.
200mV range Maximum Input: 250V dc or 250V ac rms.
Maximum Input: 1000V dc or 700V ac rms.

AC Voltage

Range

Resolution

Accuracy

2.000V

1mV

+1.0% of rdg + 3 dgts

20.00V

10mV

200.0V

100mV

700V

1V

+1.2% of rdg + 5 dgts

Input Impedance: 10M.
Frequency Range:50 to 400Hz 
Maximum Input: 1000V dc or 700V ac rms.

DC Current

Range

Resolution

Accuracy

2.000mA

1µA

+1.0% of rdg + 3 dgts

20.00mA

10uA

+1.5% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+2.5% of rdg + 10 dgts

DC Current 

  7 

  8 

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

AC Current 

Range

Resolution Accuracy

2.000mA

1µA

+1.2% of rdg + 3 dgts

20.00mA

10µA

+2.0% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+3% of rdg + 10 dgts

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

Resistance

Range

Resolution

Accuracy

200.0

Ω

0.1

Ω

+1.0% of rdg + 4 dgts

2.000k

Ω

1

Ω

+1.0% of rdg + 2 dgts

20.00k

Ω

10

Ω

+1.2% of rdg + 2 dgts

200.0k

Ω

100

Ω

2.000M

Ω

1k

Ω

20.00M

Ω

10k

Ω

+2.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Capacitance

   

Range

Resolution Accuracy

2.000nF

1pF

+4.0% of rdg+10dgts

20.00nF

10pF

+4.0% of rdg + 3 dgts

200.0nF

0.1nF

+4.0% of rdg + 3 dgts

2.000µF

1nF

200.0µF

0.1µF

+4.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Frequency

Range

Resolution Accuracy

20.00kHz

0.01Hz

+1.5% of rdg + 5 dgts

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V dc or ac rms.

Temperature

Range

Resolution Accuracy

-20oC~+1300oC

1 oC

+3% of rdg +3dgts

-4 oF~+2000 oF

1oF

Sensor: Type K Thermocouple
Input Protection: 0.5A/250V Fuse.

Diode Test

Test current: 1mA typical
Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than50 Test current: <0.3mA
Overload protection: 250V dc or ac rms.

OPERATION

WARNING

: Risk of electrocution. High-voltage 

circuits, both AC and DC, are very dangerous 

and should be measured with great care.

If “1” appears in the display during a measurement, the value 
exceeds the range you have 

selected. Change to a higher range. 

NOTE

: On some low AC and DC voltage ranges, with the test leads 

  7 

  8 

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

AC Current 

Range

Resolution Accuracy

2.000mA

1µA

+1.2% of rdg + 3 dgts

20.00mA

10µA

+2.0% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+3% of rdg + 10 dgts

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

Resistance

Range

Resolution

Accuracy

200.0

Ω

0.1

Ω

+1.0% of rdg + 4 dgts

2.000k

Ω

1

Ω

+1.0% of rdg + 2 dgts

20.00k

Ω

10

Ω

+1.2% of rdg + 2 dgts

200.0k

Ω

100

Ω

2.000M

Ω

1k

Ω

20.00M

Ω

10k

Ω

+2.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Capacitance

   

Range

Resolution Accuracy

2.000nF

1pF

+4.0% of rdg+10dgts

20.00nF

10pF

+4.0% of rdg + 3 dgts

200.0nF

0.1nF

+4.0% of rdg + 3 dgts

2.000µF

1nF

200.0µF

0.1µF

+4.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Frequency

Range

Resolution Accuracy

20.00kHz

0.01Hz

+1.5% of rdg + 5 dgts

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V dc or ac rms.

Temperature

Range

Resolution Accuracy

-20oC~+1300oC

1 oC

+3% of rdg +3dgts

-4 oF~+2000 oF

1oF

Sensor: Type K Thermocouple
Input Protection: 0.5A/250V Fuse.

Diode Test

Test current: 1mA typical
Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than50 Test current: <0.3mA
Overload protection: 250V dc or ac rms.

OPERATION

WARNING

: Risk of electrocution. High-voltage 

circuits, both AC and DC, are very dangerous 

and should be measured with great care.

If “1” appears in the display during a measurement, the value 
exceeds the range you have 

selected. Change to a higher range. 

NOTE

: On some low AC and DC voltage ranges, with the test leads 

  7 

  8 

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

AC Current 

Range

Resolution Accuracy

2.000mA

1µA

+1.2% of rdg + 3 dgts

20.00mA

10µA

+2.0% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+3% of rdg + 10 dgts

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

Resistance

Range

Resolution

Accuracy

200.0

Ω

0.1

Ω

+1.0% of rdg + 4 dgts

2.000k

Ω

1

Ω

+1.0% of rdg + 2 dgts

20.00k

Ω

10

Ω

+1.2% of rdg + 2 dgts

200.0k

Ω

100

Ω

2.000M

Ω

1k

Ω

20.00M

Ω

10k

Ω

+2.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Capacitance

   

Range

Resolution Accuracy

2.000nF

1pF

+4.0% of rdg+10dgts

20.00nF

10pF

+4.0% of rdg + 3 dgts

200.0nF

0.1nF

+4.0% of rdg + 3 dgts

2.000µF

1nF

200.0µF

0.1µF

+4.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Frequency

Range

Resolution Accuracy

20.00kHz

0.01Hz

+1.5% of rdg + 5 dgts

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V dc or ac rms.

Temperature

Range

Resolution Accuracy

-20oC~+1300oC

1 oC

+3% of rdg +3dgts

-4 oF~+2000 oF

1oF

Sensor: Type K Thermocouple
Input Protection: 0.5A/250V Fuse.

Diode Test

Test current: 1mA typical
Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than50 Test current: <0.3mA
Overload protection: 250V dc or ac rms.

OPERATION

WARNING

: Risk of electrocution. High-voltage 

circuits, both AC and DC, are very dangerous 

and should be measured with great care.

If “1” appears in the display during a measurement, the value 
exceeds the range you have 

selected. Change to a higher range. 

NOTE

: On some low AC and DC voltage ranges, with the test leads 

AC Current 

  7 

  8 

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

AC Current 

Range

Resolution Accuracy

2.000mA

1µA

+1.2% of rdg + 3 dgts

20.00mA

10µA

+2.0% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+3% of rdg + 10 dgts

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

Resistance

Range

Resolution

Accuracy

200.0

Ω

0.1

Ω

+1.0% of rdg + 4 dgts

2.000k

Ω

1

Ω

+1.0% of rdg + 2 dgts

20.00k

Ω

10

Ω

+1.2% of rdg + 2 dgts

200.0k

Ω

100

Ω

2.000M

Ω

1k

Ω

20.00M

Ω

10k

Ω

+2.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Capacitance

   

Range

Resolution Accuracy

2.000nF

1pF

+4.0% of rdg+10dgts

20.00nF

10pF

+4.0% of rdg + 3 dgts

200.0nF

0.1nF

+4.0% of rdg + 3 dgts

2.000µF

1nF

200.0µF

0.1µF

+4.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Frequency

Range

Resolution Accuracy

20.00kHz

0.01Hz

+1.5% of rdg + 5 dgts

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V dc or ac rms.

Temperature

Range

Resolution Accuracy

-20oC~+1300oC

1 oC

+3% of rdg +3dgts

-4 oF~+2000 oF

1oF

Sensor: Type K Thermocouple
Input Protection: 0.5A/250V Fuse.

Diode Test

Test current: 1mA typical
Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than50 Test current: <0.3mA
Overload protection: 250V dc or ac rms.

OPERATION

WARNING

: Risk of electrocution. High-voltage 

circuits, both AC and DC, are very dangerous 

and should be measured with great care.

If “1” appears in the display during a measurement, the value 
exceeds the range you have 

selected. Change to a higher range. 

NOTE

: On some low AC and DC voltage ranges, with the test leads 

  7 

  8 

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

AC Current 

Range

Resolution Accuracy

2.000mA

1µA

+1.2% of rdg + 3 dgts

20.00mA

10µA

+2.0% of rdg + 3 dgts

200.0mA

100µA

20.00A

10mA

+3% of rdg + 10 dgts

Overload Protection: 0.2A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 400 Hz
Maximum Input: 200mA dc or 200mA ac rms on mA ranges, 
20A dc or ac rms on 20A range.

Resistance

Range

Resolution

Accuracy

200.0

Ω

0.1

Ω

+1.0% of rdg + 4 dgts

2.000k

Ω

1

Ω

+1.0% of rdg + 2 dgts

20.00k

Ω

10

Ω

+1.2% of rdg + 2 dgts

200.0k

Ω

100

Ω

2.000M

Ω

1k

Ω

20.00M

Ω

10k

Ω

+2.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Capacitance

   

Range

Resolution Accuracy

2.000nF

1pF

+4.0% of rdg+10dgts

20.00nF

10pF

+4.0% of rdg + 3 dgts

200.0nF

0.1nF

+4.0% of rdg + 3 dgts

2.000µF

1nF

200.0µF

0.1µF

+4.0% of rdg + 5 dgts

Input Protection: 250V dc or 250V ac rms.

Frequency

Range

Resolution Accuracy

20.00kHz

0.01Hz

+1.5% of rdg + 5 dgts

Sensitivity: 200mV~ 10V RMS ;
Overload protection: 250V dc or ac rms.

Temperature

Range

Resolution Accuracy

-20oC~+1300oC

1 oC

+3% of rdg +3dgts

-4 oF~+2000 oF

1oF

Sensor: Type K Thermocouple
Input Protection: 0.5A/250V Fuse.

Diode Test

Test current: 1mA typical
Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than50 Test current: <0.3mA
Overload protection: 250V dc or ac rms.

OPERATION

WARNING

: Risk of electrocution. High-voltage 

circuits, both AC and DC, are very dangerous 

and should be measured with great care.

If “1” appears in the display during a measurement, the value 
exceeds the range you have 

selected. Change to a higher range. 

NOTE

: On some low AC and DC voltage ranges, with the test leads 

Frequency

Temperature  

Содержание OXD-516-4320M

Страница 1: ...well co uk Terminal blocking prevention Temperature measurement Extra large display Hold function Backlight Tilt stand Voltage DC 200mV 1000V Voltage AC 2V 700V Current DC 2mA 20A Current AC 2mA 20A R...

Страница 2: ...6 F 9 200m 200 2 TEMP MAX 200mA FUSED mA 20 20 A F V C C 20 COM F V 2M 2 20m 2K 200 200 20K 200m V 1000 200 2 20 700 Hz 20K 20M HOLD 10 F 2 C 20 HOLD 200 X A M C F 4 3 2 8 1 7 Input Limits Function Ma...

Страница 3: ...ndicates a potentially hazardous situation which if not avoided could result in death or serious injury This CAUTION symbol indicates a potentially hazardous situation which if not avoided may result...

Страница 4: ...0 1mV 0 5 of rdg 2 dgts 2 000V 1mV 20 00V 10mV 200 0V 100mV 1000V 1V 0 8 of rdg 2 dgts Input Impedance 10M 200mV range Maximum Input 250V dc or 250V ac rms Maximum Input 1000V dc or 700V ac rms AC Vo...

Страница 5: ...tection 250V dc or 250V ac rms Frequency Range Resolution Accuracy 20 00kHz 0 01Hz 1 5 of rdg 5 dgts Sensitivity 200mV 10V RMS Overload protection 250V dc or ac rms Temperature Range Resolution Accura...

Страница 6: ...the Multimeter The indicators HOLD will appear in the display 2 Press the DATA HOLD button to return to normal operation BACK LIGHT BUTTON The BACK LIGHT BUTTON is used to turn the back light on only...

Страница 7: ...ent measurements up to 20A DC set the function switch to the A position and insert the red test lead banana plug into the 20A jack 4 Remove power from the circuit under test then open up the circuit a...

Страница 8: ...ately 50 an audible signal will sound The display will also show the actual resistance DIODE TEST WARNING To avoid electric shock do not test any diode that has voltage on it 1 Set the function switch...

Страница 9: ...r making sure to observe the correct polarity 4 Switch the multimeter on and select C or F 5 Touch the temperature probe head to the part whose temperature you wish to measure Keep the probe touching...

Страница 10: ...s using a Phillips head screwdriver 3 Insert the battery into battery holder observing the correct polarity 4 Put the battery cover back in place Secure the two screws WARNING To avoid electric shock...

Страница 11: ...he back cover using a Phillips head screwdriver 3 Remove the old fuse from its holder by gently pulling it out 4 Install the new fuse into the holder 5 Always use a fuse of the proper size and value 0...

Страница 12: ...ical goods that have reached the end of their life must be collected separately and returned to an environmentally compatible recycling facility Do not dispose of electrical goods with domestic waste...

Отзывы: