Safety
The following general safety precautions must be observed during the operation, service
and repair of this instrument.
2.1
Protective Ground
To minimise shock hazard the instrument must be connected to an electrical ground. The
ground wire (green/yellow) in the instrument power cable must be connected to the
installation electrical ground system. Do not use extension cords without a protective
earth conductor. Do not disconnect the protective ground inside or outside the
instrument. Do not have external circuits connected to the instrument when its
protective ground is disconnected.
2.2
Working Environment
The ILM indicator unit is not designed to be water or splash proof. Therefore it should
not be exposed to rain or excessive moisture.
Neither the indicator unit nor the probes are designed to be used in areas where there
are flammable or explosive gases or fumes.
2.3
Repair and Adjustment
Some internal adjustments can be made to ILM. Although we do not encourage you to
make these adjustments we try to supply you with enough information to allow you to
do it safely. Disconnect the AC power supply before you remove the covers or fuses,
because dangerous voltages are accessible on the circuit board and other components. It
is not sufficient to switch off the main power switch. Capacitors inside the instrument
and the power connector filter, may remain charged after removal of AC power.
Discharge these carefully before you start work.
Some fault finding operations can only be carried out with power connected to the
instrument. If you have to reconnect the AC power supply with the protective covers
removed you must remember that you are putting your life at risk. You should only do
this type of work if you are suitably qualified and sufficiently skilled to understand all the
risks you are taking.