![Oxford Instruments Andor Sona 2.0B-11 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/oxford-instruments/andor-sona-2-0b-11/andor-sona-2-0b-11_hardware-manual_1664924047.webp)
version 1.5 rev 4 Dec 2019
47
Sona &
Marana
External Power Supply Requirements
Low Voltage Supply Input
15 V +/- 5%
Low Voltage Supply Current
8 A
Low Voltage Supply Cable Connector
Right-angle Plug: Fischer WSO 104 A037-130+
Straight Plug: Fischer S 104 A037-130+
Required Cable Clamp Set: Fischer E3 104.3/6.7+B
Low Voltage Supply Connector Pin Connections
Pins 1 & 2: +15 V
Pins 3 & 4: 0 V
Shield: 0 V
Low Voltage Supply Cable Connector
Solder-side View
1
2
3
4
S(-)
(-)
(+)
Ripple
150 mV peak-to-peak
In-rush Current Capability
Shall start up a load whose in-rush current from a 0.15 Ω source resistance is 2.7 A min.
peak and a pulse width of 15 ms min. measured at half the peak
Safety
Certified to an appropriate IEC standard, e.g. IEC 60950-1, and meet the reinforced
insulation from mains requirement of IEC 61010-1
Environmental
Ensure that the EPS meets the environmental specification of the overall product (see
above)
Camera Power Specifications
Mains Input for Supplied External Power Supply
100-240 VAC, 50-60 Hz
Power Consumption
External Power Supply: Air cooling of sensor to -25°C: 40 W typical/ 114 W
max
Water cooling of sensor to -45°C: 40 W typical/ 114 W max
Camera Only: Air cooling of sensor to -25°C: 35 W typical / 100W max
Water cooling of sensor to -45°C: 35 W typical/ 100 W max
Voltage Rating
15 V
Current Rating
8 A