4.Junior User Guidebook
17
Figure 4-12 Trigger Control Zone
1.
Press the "
Trigger Menu
" button and call out the trigger menu. With the
operations of the menu selection buttons, the trigger setting can be changed.
2.
Use the "
TRIG
LEVEL
"
knob to change the trigger level setting.
By rotating the "
TRIG LEVEL
" knob, the trigger indicator in the screen will
move up and down. With the movement of the trigger indicator, it can be
observed that the trigger level value displayed in the screen changes accordingly.
PS
: Turning the
TRIG LEVEL
knob can change trigger level value and it is also
the hotkey to set trigger level back to 0.
3.
Press the button "
50%
" to set the trigger level as the vertical midpoint values of
the amplitude of the trigger signal.
4.
Press the "
Force
" button to force a trigger signal, which is mainly applied to the
"Normal" and "Single" trigger modes.
Introduction to the Touchscreen Controls
When
Gesture
is ON, which is set in menu
Utility
->
Display
->
Gesture
(
Page 45