4.Advanced User Guidebook
44
when
Polarity
Choose the polarity
Select pulse width condition and adjust the
M
knob
or click
to set time, press
panel
button or click
to move cursor to choose
which digit to be set.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Sensitivity
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
knob or click
to set time
interval before another trigger occur, press
panel button or click
to move cursor to
choose which digit to be set.
Set channel sensitivity(Range: 0.3div-10div).
5.Runt Trigger
Trigger pulses that pass through one trigger level but fail to pass through the other
trigger level. Shown as below figure,
In Runt Trigger mode, the trigger setting information is displayed on bottom right of
the screen, for example,
,
indicates that trigger type is runt,
trigger source is CH1, polarity is positive, 0.00mV is the differential between up level
and low level threshold.
Runt Trigger
Runt Trigger menu
list:
MENU
SETTING
INSTRUCTION