background image

8.Communication with Android Device via Wi-Fi (Optional)

 

 

124 

     

 

(6) 

Select 

Save set 

in the bottom menu to save current settings.   

(7) 

In  Android  device,  enter  the  Wi-Fi  settings,  select  the  oscilloscope  access  point  to 

connect, enter the set password. 

 

(8) 

In Android device, launch the application. 

(9) 

Click 

connect

 on the left. 

 

(10) 

Set 

IP

 and 

Port

 to the same value as the setting in the oscilloscope. Click 

Confirm

 

to connect. Click 

History

 to recall the history settings.   

Содержание Dual-Channel Series

Страница 1: ...XDS3000 Dual Channel Series Digital Storage Oscilloscopes User Manual www owon com ...

Страница 2: ...eplace all materials published The information in this manual was correct at the time of printing However LILLIPUT will continue to improve products and reserves the rights to change specification at any time without notice is the registered trademark of the LILLIPUT Company Fujian LILLIPUT Optoelectronics Technology Co Ltd No 19 Heming Road Lantian Industrial Zone Zhangzhou 363005 P R China Tel 8...

Страница 3: ...signated service centre a copy of the customers proof of purchase is also required This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care We shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than our company representatives to install repair or serv...

Страница 4: ...tical System 14 Introduction to the Horizontal System 15 Introduction to the Trigger System 16 Touchscreen Controls Touchscreen is optional 17 Operate the Menu through Touchscreen 17 Gestures in Normal Mode 18 Gestures in Wave Zoom Mode 21 Other Operations Using Touchscreen 23 Magnifier Function Only for XDS3102AP XDS3202A with touchscreen 23 4 Advanced User Guidebook 26 How to Set the Vertical Sy...

Страница 5: ... Use Executive Buttons 96 How to Print the Screen Image 97 5 Use theArbitrary Function Generator Optional 99 Output Connection 99 To Set Channels 99 To Set Signals 100 To Output Sine Signals 100 To Set the Frequency 100 To Set the Period 101 To Set the Start Phase 101 To Set the Amplitude 101 To Set the Offset 102 To Set the High Level 102 To Set the Low Level 102 To Output Square Signals 102 To O...

Страница 6: ... router 117 Using Wi Fi to Connect with PC Optional 119 Connect with PC as Wi Fi Access Point 119 Connect with PC as Wi Fi Station 120 8 CommunicationwithAndroidDeviceviaWi Fi Optional 123 How to Connect 123 Connect with APP as Wi Fi Access Point 123 Connect with APP as Wi Fi Station 125 User Interface 127 Gestures Control 129 9 Demonstration 131 Example 1 Measurement a Simple Signal 131 Example 2...

Страница 7: ...veform Generator Optional 146 Multimeter Optional 147 General Technical Specifications 148 12 Appendix 149 Appendix A Enclosure 149 Appendix B General Care and Cleaning 149 Appendix C Battery Using Guide 150 ...

Страница 8: ...output terminal When powered by AC power it is not allowed to measure AC power source directly because the testing ground and power cord ground conductor are connected together otherwise it will cause short circuit When powered by battery the product must ground connection To avoid electric shock there must be a ground wire connect between ground and the ground port on the back of product panel Ch...

Страница 9: ... accessible potentially Caution It indicates a potential damage to the instrument or other property might occur Safety Symbols Symbols on the product The following symbol may appear on the product Hazardous Voltage Refer to Manual Protective Earth Terminal Chassis Ground Test Ground To avoid body damage and prevent product and connected equipment damage carefully read the following safety informat...

Страница 10: ...h the ports Warning To avoid fire or electrical shock when the oscilloscope input signal connected is more than 42V peak 30Vrms or on circuits of more than 4800VA please take note of below items Only use accessory insulated voltage probes and test lead Check the accessories such as probe before use and replace it if there are any damages Remove probes test leads and other accessories immediately a...

Страница 11: ...rface How to implement the general inspection How to implement the function inspection How to make a probe compensation How to set the probe attenuation coefficient How to use the probe safely How to implement an auto calibration Introduction to the vertical system Introduction to the horizontal system Introduction to the trigger system Touchscreen Controls Touchscreen is optional ...

Страница 12: ...ch you can set the different options for the current menu The other buttons are function buttons through which you can enter different function menus or obtain a specific function application directly 8 2 5 3 7 4 6 1 Figure 3 1 Front panel 1 Display area 2 Control button and knob area 3 Probe Compensation Measurement signal 5V 1kHz output 4 EXT Trigger Input 5 Signal Input Channel 6 Copy button Yo...

Страница 13: ...l Menu Buttons Select the right menu item Select the bottom menu item Remove the left and right menu Figure 3 2 Menu Buttons Rear Panel 3 10 9 8 7 13 12 11 1 2 6 4 5 Figure 3 3 Rear Panel 1 Handle 2 Air vents 3 Input terminals of multimeter optional 4 AC power input jack 5 Fuse 6 Foot stool Adjust the tilt angle of the oscilloscope 7 VGA port To connect the oscilloscope with a monitor or a project...

Страница 14: ...nerator The output type can be set on the menu Utility menu Output Output 13 Out 1 port Output single channel or CH1 Output dual channel of optional waveform generator Control Area 8 7 6 1 2 3 11 10 5 4 9 Figure 3 4 Control Area Overview 1 Function button area Total 11 buttons 2 Waveform generator controls optional or DAQ Multimeter Recorder see Multimeter Recorder on P112 P F Pass Fail see Pass F...

Страница 15: ...nal function 9 DMM optional Multimeter or Snapshot Shortcut button for measurement snapshot 10 Direction key Move the cursor of the focused parameter 11 M knob Multipurpose knob when a symbol appears on the menu it indicates you can turn the M knob to select the menu or set the value You can push it to close the menu on the left and right User Interface Introduction 1 3 10 21 16 29 26 8 9 11 13 18...

Страница 16: ... pointer shows the trigger level position 17 The waveform of CH1 18 The two blue dotted lines indicate the horizontal position of cursor measurement 19 The waveform of CH2 20 It indicates the measured type and value of the corresponding channel T means period F means frequency V means the average value Vp the peak peak value Vr the root mean square value Ma the maximum amplitude value Mi the minim...

Страница 17: ...ording to the following steps 1 Check whether there is any damage caused by transportation If it is found that the packaging carton or the foamed plastic protection cushion has suffered serious damage do not throw it away first till the complete device and its accessories succeed in the electrical and mechanical property tests 2 Check the Accessories The supplied accessories have been already desc...

Страница 18: ...nu select Default in the bottom menu The default attenuation coefficient set value of the probe in the menu is 10X 2 Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel Align the slot in the probe with the plug in the CH1 connector BNC and then tighten the probe with rotating it to the right side Connect the probe tip and the ground clamp to the connector ...

Страница 19: ...tip with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector and then push the Autoset button on the front panel 2 Check the displayed waveforms and regulate the probe till a correct compensation is achieved see Figure 3 7 and Figure 3 8 Overcompensated Compensated correctly Under compensated Figure 3 7 Displayed ...

Страница 20: ... in 5MHz To use the full bandwidth of the oscilloscope the switch must be set to 10X Identify the Probe Attenuation Coefficient Automatically The oscilloscope can identify the probe attenuation coefficient of the 100 1 impedance 5K 20 or 10 1 impedance 10K 20 probe with the identifying pin When you attach the probe the oscilloscope set the attenuation automatically on the oscilloscope vertical men...

Страница 21: ... time This program must be executed whenever the change of ambient temperature is 5 or over Before performing a self calibration disconnect all probes or wires from the input connector Push the Utility button select Function in the bottom menu select Adjust in the left menu select Self Cal in the bottom menu run the program after everything is ready Introduction to the Vertical System As shown in ...

Страница 22: ...hen the trace position is far out of the screen and want it to get back to the screen center immediately 2 Change the Vertical Setting and Observe the Consequent State Information Change With the information displayed in the status bar at the bottom of the waveform window you can determine any changes in the channel vertical scale factor Turn the Vertical Scale knob and change the Vertical Scale F...

Страница 23: ...o set the triggering displacement back to 0 as a shortcut key 3 Push the Horizontal HOR button to switch between the normal mode and the wave zoom mode Introduction to the Trigger System As shown in Figure 3 13 there are one knob and three buttons make up Trigger Controls The following practices will direct you to be familiar with the setting of the trigger system gradually Figure 3 13 Trigger Con...

Страница 24: ...gh Touchscreen Select a menu item Touch the menu items in the bottom menu or in the right menu or in the left menu Switch menu items If there are options that can be switched in the menu you can repeatedly touch the area of the menu item to switch or push the corresponding button to switch See figure below Press repeatedly to switch the options Scroll the list If there is a scroll bar in the left ...

Страница 25: ...t in selected state Set the vertical position of the selected channel Vertical Position knob Swipe up or down on the screen Set the trigger level of the source in trigger menu Trigger Level knob Swipe up or down on the area nearby the trigger pointer on the right Set the horizontal position Horizontal Position knob Swipe left or right on the display area ...

Страница 26: ...Double Zoom and Single Zoom In touchable menu pane if DoubleZoom is selected in the display area pinch and spread horizontally to change the time base pinch and spread vertically to change the voltage division of current channel ...

Страница 27: ...uidebook 20 In touchable menu pane if SingleZoom is selected click anywhere on the Display area to show the touching control panel Set the voltage division Vertical Scale knob Click on the left upper area to increase ...

Страница 28: ...rease the voltage division of CH2 Set the horizontal time base Horizontal Scale knob Click on the left area to increase the voltage division click on the right area to decrease the voltage division Gestures in Wave Zoom Mode Push the Horizontal HOR button to enter into zoom mode The top half of the display shows the Main window and the bottom half displays the Zoom window The Zoom window is a magn...

Страница 29: ...3 Junior User Guidebook 22 Double Zoom Single Zoom ...

Страница 30: ...lue in the menu item Click to increase the value of cursor position Move the cursor Click to decrease the value of cursor position Magnifier Function Only for XDS3102AP XDS3202A with touchscreen The magnifier window can display the magnified wave of the wave selection when the magnifier function is enabled Click the icon on the left top of the display area the magnifier window and menu will be sho...

Страница 31: ... Vertical 1 2 4 8 16 32 Vertical magnification factor The size of the magnifier window is fixed Therefor the greater this value the shorter the vertical height of the wave selection Transparent ON OFF Make the magnifier window transparent or not The Cursor Measurement in Magnifier Window When the magnifier function is enabled push Cursor panel key to display the cursor menu In the bottom menu sele...

Страница 32: ...3 Junior User Guidebook 25 Note The magnifier function is disabled in wave zoom mode waveform math FFT XY mode slow scan mode stop state average acquire mode and persist mode ...

Страница 33: ...ll deal with the following topics mainly How to Set the Vertical System How to Set the Horizontal System How to Set the Trigger Decoding System How to Implement the Sampling Setup How to Set the Display System How to Save and Recall Waveform How to Record Playback Waveforms How to Clone and Recall a waveform How to Implement the Auxiliary System Function Setting How to Update your Instrument Firmw...

Страница 34: ...on of the Channel Menu is shown as the following list Function Menu Setting Description Coupling DC AC GROUND Pass both AC and DC components of the input signal Block the DC component of the input signal Disconnect the input signal Inverted ON OFF Display inverted waveform Display original waveform Probe Attenu 0 001X to 1000X Step by 1 2 5 Match this to the probe attenuation factor to have an acc...

Страница 35: ...M knob to set it as 10 3 To measure current by probing the voltage drop across a resistor Take the Channel 1 as an example if you are measuring current by probing the voltage drop across a 1Ω resistor the operation steps are shown as follows 1 Push the CH1 button to show CH1 SETUP menu 2 Select Probe in the bottom menu In the right menu set MeasCurr as YES the A V radio menu will appear below Sele...

Страница 36: ...example the operation steps are as follows 1 Press the CH1 button to call up the CH1 setting menu 2 In the bottom menu select More in the right menu select the desired Unit The default unit is V Use Mathematical Manipulation Function The Mathematical Manipulation function is used to show the results of the addition multiplication division and subtraction operations between two channels the FFT ope...

Страница 37: ...on Intg Diff Sqrt and user defined function DIR channel CH1 CH2 Select channel type low pass Only the signals whose frequencies are lower than the current cut off frequency can pass the filter high pass Only the signals whose frequencies are greater than the current cutoff frequency can pass the filter band pass Only the signals whose frequencies are greater than the cutoff frequency down and lowe...

Страница 38: ...ttom menu 3 In the right menu select Factor1 as CH1 4 Select Sign as in the right menu 5 In the right menu select Factor2 as CH2 6 Select Vertical div in the right menu turn the M knob to adjust the vertical position of Math waveform 7 Select Vertical V div in the right menu turn the M knob to adjust the vertical division of Math waveform User defined function 1 Press the Math button to display th...

Страница 39: ...he input signal on Oscilloscope You can match these frequencies with known system frequencies such as system clocks oscillators or power supplies FFT function in this oscilloscope transforms 8192 data points of the time domain signal into its frequency components mathematically the record length should be 10K or above The final frequency contains 4096 points ranging from 0Hz to Nyquist frequency T...

Страница 40: ...he event are significantly different Rectangle Best solution for frequency worst for magnitude Best type for measuring the frequency spectrum of nonrepetitive signals and measuring frequency components near DC Recommend to use for Transients or bursts the signal level before and after the event are nearly equal Equal amplitude sine waves with frequencies those are very close Broadband random noise...

Страница 41: ...ing amplitudes the side lobe level and shape factor are closest to the traditional Gaussian RBW This window is also good for random signals Bartlett The Bartlett window is a slightly narrower variant of the triangular window with zero weight at both ends Figure 4 1 Figure 4 2 Figure 4 3 Figure 4 4 Figure 4 5 Figure 4 6 are examples for measuring sine wave with a frequency of 1kHz under the selecti...

Страница 42: ...4 Advanced User Guidebook 35 Figure 4 2 Rectangle window Figure 4 3 Blackman window ...

Страница 43: ...4 Advanced User Guidebook 36 Figure 4 4 Hanning window Figure 4 5 Kaiser window ...

Страница 44: ...t is Nyquist frequency The Nyquist frequency is the highest frequency that any real time digitizing oscilloscope can acquire without aliasing This frequency is half of the sample rate Frequencies above the Nyquist frequency will be under sampled which causes aliasing So pay more attention to the relation between the frequency being sampled and measured Use Vertical Position and Scale Knobs 1 The V...

Страница 45: ... of all channels include those obtained from the mathematical manipulation the analytic resolution of which changes with the time base 2 Horizontal Scale knob it is used to set the horizontal scale factor for setting the main time base or the window 3 Horizontal HOR button push it to switch between the normal mode and the wave zoom mode For more detailed operations see the introductions below Zoom...

Страница 46: ...s to acquire data it will collect enough data to draw waveform on left of trigger point DSO continues to acquire data while waiting for trigger condition to occur Once it detects a trigger it will acquire enough data continuously to draw the waveform on right of trigger point Trigger control area consists of 1 knob and 2 menu buttons Trigger Level The knob that set the trigger level push the knob ...

Страница 47: ...pulse trigger runt trigger windows trigger timeout trigger and Nth edge trigger Edge Trigger It occurs when the trigger input passes through a specified voltage level with the specified slope Video Trigger Trigger on fields or lines for standard video signal Slope Trigger The oscilloscope begins to trigger according to the signal rising or falling speed Pulse Trigger Find pulses with certain width...

Страница 48: ...onent Allow all component pass Block the high frequency signal only low frequency component pass Block the low frequency signal only high frequency component pass LF menu is only for certain models Slope Trigger on rising edge Trigger on falling edge Mode Holdoff Auto Normal Single Holdoff Sensitivity Acquire waveform even no trigger occurs Acquire waveform when trigger occurs When trigger occurs ...

Страница 49: ... line Synchronic trigger in video field Synchronic trigger in video odd filed Synchronic trigger in video even field Synchronic trigger in designed video line turn the M knob or click to set the line number Mode Holdoff Auto Acquire waveform even no trigger occurred 3 Slope Trigger Slope trigger sets the oscilloscope as the positive negative slope trigger within the specified time In Slope Trigger...

Страница 50: ...or click to set time interval before another trigger occur press panel button or click to move cursor to choose which digit to be set Set Holdoff time as 100 ns 4 Pulse Width Trigger Pulse trigger occurs according to the width of pulse The abnormal signals can be detected through setting up the pulse width condition In Pulse Width Trigger mode the trigger setting information is displayed on bottom...

Страница 51: ...lick to set time interval before another trigger occur press panel button or click to move cursor to choose which digit to be set Set channel sensitivity Range 0 3div 10div 5 Runt Trigger Trigger pulses that pass through one trigger level but fail to pass through the other trigger level Shown as below figure In Runt Trigger mode the trigger setting information is displayed on bottom right of the s...

Страница 52: ...h digit to be set Trigger when runt pulse is greater than the set pulse width Trigger when runt pulse equals to the set pulse width Trigger when runt pulse is lower than the set pulse width Mode Holdoff Auto Normal Single Holdoff Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform then stop 100 ns 10 s adjust M knob or click to ...

Страница 53: ...on the positive Windows pulse Negative Polarity the oscilloscope triggers on the negative Windows pulse Enter Triggers when the trigger signal enters the specified trigger level range Exit Triggers when the trigger signal exits the specified trigger level range Time Specify the hold time of the input signal after entering the specified trigger level The oscilloscope triggers when the accumulated h...

Страница 54: ...igger level Start timing when the falling edge of the input signal passes through the trigger level Configure Idle Time Set idle time Idle time means the minimum time of idle clock before searching data that can meet trigger conditions Available range is 30ns 10s default 100ns Mode Holdoff Auto Normal Single Holdoff Sensitivity Acquire waveform even no trigger occurred Acquire waveform when trigge...

Страница 55: ... Source CH1 CH2 Select CH1 as the trigger source Select CH2 as the trigger source Edge Edge Trigger on the rising edge of the input signal when voltage level meets the specified trigger level Trigger on the falling edge of the input signal when voltage level meets the specified trigger level Configure Idle Time Set idle time before the edge counting in Nth Edge Trigger Adjust M knob or click to se...

Страница 56: ...menu list Menu Settings Instruction Alternate Mode Edge Set vertical channel trigger type as edge trigger Source CH1 CH2 Channel 1 as trigger signal Channel 2 as trigger signal Coupling AC DC Block the direct current component Allow all component pass Slope Trigger on rising edge Trigger on falling edge Mode Holdoff Auto Holdoff Sensitivity Acquire waveform even no trigger occurs 100 ns 10 s turn ...

Страница 57: ... is more than the set time Trigger when the time of true condition is lower than the set time Trigger when the time of true condition is equal to the set time Goes False Is True Is True Is True Mode Holdoff Auto Normal Single Holdoff Sensitivity Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs acquire one waveform then stop 100 ns 10 s adjust M k...

Страница 58: ...hoosing this condition press Configure to enter detailed settings Error Trigger when error frame is detected After choosing this condition press Configure to enter detailed settings Chk Error Trigger when Chk Error is detected After choosing this condition press Configure to enter detailed settings Data Trigger on the last bit of the preset data After choosing this condition press Configure to ent...

Страница 59: ...The I2C serial bus consists of SCL and SDA The transmission rate is determined by SCL and the transmission data is determined by SDA As shown in below figure oscilloscope can trigger on the start restart stop ack lost specific device address or data value also device address and data value at the same time In I2C bus trigger mode the trigger setting information is displayed on bottom right of the ...

Страница 60: ...atF orma t Byte lengt h Set data byte length available range 1 5 bytes Adjust M knob or click to set byte length Select the data bit ranges from 0 to byte length 8 1 Set data to be H L or X H or L Set all the data bits to be the specified value in Data Curr entB it Data All Bits Addr Data Trigger when Address and Data conditions are met at the same time Mode Holdoff Auto Normal Single Acquire wave...

Страница 61: ...M knob or click to set Data Bits Set the number of the data bits ranges from 0 31 adjust M knob or click to set Current Bit Set the value of the current data bit as H L or X H or L Set all the data bits to be the specified value in Data Data Bits Current Bit Data All Bits Mode Holdoff Auto Normal Single Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger o...

Страница 62: ...me from the start of the bit s time to the sample point time in the bit s time The range is 5 to 95 Common Baud Turn the M knob to select from the Baud list on the left Custom Baud Turn the M knob or tap on in touchscreen to set the Baud The range is 10 000 to 1 000 000 Tip You can select the nearest value in Common Baud and then adjust it in this menu Condition Start Trigger on the start frame of...

Страница 63: ... In trigger menu select Bus trigger and select bus type as RS232 set parameters based on the characteristics of the signal trigger the signal correctly and obtain stable display Refer to RS232 Trigger on page 50 4 Push the Decode button on the front panel Select bus type as RS232 set parameters based on the characteristics of the signal When the parameters are set correctly the information carried...

Страница 64: ...rage device 2 I2C Decoding To decode I2C signal 1 Connect the clock line SCLK and the data line SDA of the I2C signal to the Signal Input Channels of the oscilloscope 2 Adjust to the proper time base and voltage division 3 In trigger menu select Bus trigger and select bus type as I2C set parameters based on the characteristics of the signal trigger the signal correctly and obtain stable display Re...

Страница 65: ...he data line SDA of the SPI signal to the Signal Input Channels of the oscilloscope 2 Adjust to the proper time base and voltage division 3 In trigger menu select Bus trigger and select bus type as SPI set parameters based on the characteristics of the signal trigger the signal correctly and obtain stable display Refer to SPI Trigger on page 53 4 Push the Decode button on the front panel Select bu...

Страница 66: ...y the event table Save EventTable If a USB storage device is currently connected to the instrument save the event table data in a csv spreadsheet formatted file on the external USB storage device 4 CAN Decoding To decode CAN signal 1 Connect the CAN signal to the Signal Input Channel of the oscilloscope 2 Adjust to the proper time base and voltage division 3 In trigger menu select Bus trigger and ...

Страница 67: ...event table Data Frame will not be identified CAN Decoding menu list MENU SETTING INSTRUCTION Bus Type CAN Set bus type of decoding as CAN Display Format Binary Decimal Hex ASCII Set the display format of the bus EventTable ON OFF Select ON to display the event table Save EventTable If a USB storage device is currently connected to the instrument save the event table data in a csv spreadsheet form...

Страница 68: ...oth single CH and double CH 10K 100K 1M 10M 20M 40M Single CH The description of PERF Mode menu only for XDS3000 A series is shown as follows Function Menu Setting Description PERF Mode 8 bit Set the vertical resolution A D 12 bit 14 bit only for XDS3102AP XDS3202A This menu is only for XDS3000 A series When the sample rate 250MS s ADC resolution is set to 12 bit by default When the sample rate 25...

Страница 69: ...on method according to the actual signal Sine x x interpolation Connect the sampled points with curved lines Linear interpolation Connect the sampled points with straight lines This method is suitable to rebuild the straight edged signals such as square or pulse wave Sine x x interpolation Linear interpolation How to Set the Display System Push the Display button and the Display menu is shown as f...

Страница 70: ...y The oscilloscope will start to accumulate acquisitions again Color Only for certain models Color temperature function uses color grading to indicate frequency of occurrence The hot colors like red yellow indicate frequently occurring events and the colder colors like blue green indicate rarely occurring events Function Menu Setting Description Type Dots Vect Only the sampling points are displaye...

Страница 71: ...Channel 1 is displayed in the horizontal axis and Channel 2 in the vertical axis the oscilloscope is set in the un triggered sample mode the data are displayed as bright spots The operations of all control knobs are as follows The Vertical Scale and the Vertical Position knobs of Channel 1 are used to set the horizontal scale and position The Vertical Scale and the Vertical Position knobs of Chann...

Страница 72: ...steps 1 Push Trigger Menu button set the trigger type to Single set the trigger mode to Edge select the signal source 2 Push the Display button 3 Select Counter as ON or OFF in the bottom menu How to Save and Recall a Waveform Push the Save button you can save the waveforms configures screen images record or clone the waveform The description of the Save Function Menu is shown as the following tab...

Страница 73: ...aved Close All Close all the waveforms stored in the object address Save Save the waveform of the source to the selected address Whatever the Type of save menu is set you can save the waveform by just pressing the Copy panel button in any user interface Select Type in the bottom menu in the right Format menu you can select the storage format Storage Internal External Save to internal storage or US...

Страница 74: ...l in the right menu for Source 6 Select Object Show in the bottom menu select 1 as object address in the left menu 7 Select Save in the bottom menu to save the waveform 8 Recalling Select Object Show in the bottom menu select 1 in the left menu In the right menu select Show as ON the waveform stored in the address will be shown the address number and relevant information will be displayed at the t...

Страница 75: ...equirements on P68 2 After the USB disk is installed push the Save panel button the save menu is displayed at the bottom of the screen 3 Select Type in the bottom menu select Image in the left menu 4 Select Save in the bottom menu an input keyboard used to edit the file name will pop up The default name is current system date and time Select the key in the keyboard to confirm USB disk Requirements...

Страница 76: ...1 or 2 red mark area choose Format And system will pop up a warning message click Yes Figure 4 10 Format the USB disk warning 5 Set File System as FAT32 Allocation unit size 4096 Check Perform a quick format to execute a quick format Click OK and then click Yes on the warning message ...

Страница 77: ...T32 with allocation unit size 4096 after formatting Use Minitool Partition Wizard to format Download URL http www partitionwizard com free partition manager html Tip There are many tools for the USB disk formatting on the market just take Minitool Partition Wizard for example here 1 Connect the USB disk to the computer 2 Open the software Minitool Partition Wizard ...

Страница 78: ...t or push keyboard F5 and information about the USB disk will display on the right side with red mark 1 and 2 Figure 4 13 Reload Disk 4 Right click 1 or 2 red mark area choose Format Figure 4 14 Choose format 5 Set File System FAT32 Cluster size 4096 Click OK Figure 4 15 Format setting ...

Страница 79: ...y Figure 4 18 Format successfully How to Record Playback Waveforms Wave Record function can record the input current wave You can set the interval between recorded frames in the range of 10 ms 10 s The max frame number reaches 1000 and you can get better analysis effect with playback and storage function The storage medium contains two kinds Internal and External When the storage medium is Interna...

Страница 80: ...nd Channel 2 will be recorded If a Channel is turned off while recording the waveform of the channel is invalid in the playback mode Playback Play back the wave recorded or saved Playback menu shows as follows Menu Setting Instruction Playback Mode FrameSet Start frame End frame Cur frame Interval Turn the M knob to select the number of start frame to playback 1 1000 Turn the M knob to select the ...

Страница 81: ...aymode select Operate as Play 10 To save the wave recorded select Mode in the bottom menu select Storage in the right menu Select FrameSet in the bottom menu to set the range of frames to store select Save in the bottom menu 11 To load the waveform from the internal memory select Load in the bottom menu then enter the Playback of the Mode to analyze the wave Note When playbacking the waveform the ...

Страница 82: ...rd in the right menu 6 Select FrameSet in the bottom menu set End frame and Interval in the right menu If you want to record wave to external infinitely select Infinity in the right menu the End frame will display 7 In the bottom menu set Refresh 8 In the bottom menu select Operate as Play Connect external device to the computer and wave_record_0 bin is the recorded file Open the software and do a...

Страница 83: ... line a Turn the M knob to move line b Two cursors are linked Turn the M knob to move the pair of cursors Set the cursors to select the entire screen automatically Clone Generator is available Clone Clone the waveform between two cursors and output it through the generator AG Output Only for dual channel AG Out1 Out2 If the source is CH1 or CH2 you can select the output channel of the generator If...

Страница 84: ... and set the record length to a smaller value 5 Select Save in the bottom menu To save the waveform to internal memory select Storage in the right menu as Internal Turn the M knob to select an object in the left menu select Save in the right menu To save the waveform onto a USB memory device insert a USB memory device into the port on the front panel If the icon appears on the top right of the scr...

Страница 85: ...he right menu To output the selected waveform through the generator directly The generator is optional 1 Push Save button 2 Select Type in the bottom menu in the left menu turn the M knob to select Clone 3 Select Source in the bottom menu select one channel CH1 CH2 or two channels CH1 CH2 in the right menu for Source 4 Select Line in the bottom menu Select the cursor and move it to select the desi...

Страница 86: ...e Meaning Value Comment VOLT Voltage division divided by 400 is ADC resolution When the source is CH1 CH2 it is CH1 voltage division 4 bytes float The value indicates voltage the unit is mV such as 200 mV 6 SAMP Parameter name Meaning Value Comment SAMP Sample rate 4 bytes float The unit is Sa s 7 ADCB Parameter name Meaning Value Comment ADCB ADC bit ADC resolution 4 bytes int 8 bit or 12 bit 8 C...

Страница 87: ...tom menu select Display in the left menu The description of Display Menu is shown as the follows Adjust Function Menu Setting Description Language Choose the display language of the operating system Set Time Display ON OFF On Off the date display Hour Min Setting Hour Minute Day Month Setting Date Month Year Setting Year KeyLock Lock all keys Unlock method push Trigger Menu button in trigger contr...

Страница 88: ...isplay at the left select Adjust If everything is ready select Self Cal in the bottom menu to enter the self calibration procedure of the instrument Probe checking To check whether probe attenuation is good The results contain three circumstances Overflow compensation Good compensation Inadequate compensation According to the checking result users can adjust probe attenuation to the best Operation...

Страница 89: ...n the bottom menu select Operate in the right menu as Start the test will begin 6 Save rule Select SaveRule in the bottom menu Select the save location in the left menu and then select Save in the right menu to save the rules which could be called up at once when need Select Load to call up the rule saved Note 1 When Pass Fail is ON if XY or FFT is ready to run then Pass Fail will be closed under ...

Страница 90: ...t output refer to How to Print the Screen Image on page 97 LAN Set Using the LAN port or Wi Fi the oscilloscope can be connected with a computer The oscilloscope also supports communications with Android based smart device via Wi Fi Refer to Communication with PC on page 115 and Communication with Android Device via Wi Fi Optional on page 123 for the operation steps Update Use the front panel USB ...

Страница 91: ... instrument do not power off the instrument or remove the USB memory device during the update process To update your instrument firmware do the following 1 Push the Utility button select Function in the bottom menu select Configure in the left menu select About in the bottom menu View the model and the currently installed firmware version 2 From a PC visit the website and check if the website offe...

Страница 92: ...m left of the screen The oscilloscopes provide 38 parameters for auto measurement including Period Frequency Mean PK PK RMS Max Min Top Base Amplitude Overshoot Preshoot Rise Time Fall Time PulseWidth PulseWidth Duty Cycle Duty Cycle Delay A B Delay A B Cycle RMS Cursor RMS Screen Duty FRR FRF FFR FFF LRR LRF LFR LFF Phase PulseCount PulseCount RiseEdgeCnt FallEdgeCnt Area and Cycle Area The Autom...

Страница 93: ...samples will be displayed at the bottom of the screen Std Dev Samples The number of standard deviation samples the number of samples to calculate the standard deviation Range 0 999 Reset Restart statistics on statistics Measure Only if the waveform channel is in the ON state the measurement can be performed The automatic measurement can not be performed in the following situation 1 On the saved wa...

Страница 94: ...oltage parameters The oscilloscopes provide automatic voltage measurements including Mean PK PK RMS Max Min Vtop Vbase Vamp OverShoot PreShoot Cycle RMS and Cursor RMS Figure 4 21 below shows a pulse with some of the voltage measurement points Figure 4 21 Mean The arithmetic mean over the entire waveform PK PK Peak to Peak Voltage RMS The true Root Mean Square voltage over the entire waveform ...

Страница 95: ...irst entire period of the waveform Cursor RMS The true Root Mean Square voltage over the range of two cursors The automatic measurement of time parameters The oscilloscopes provide time parameters auto measurements include Period Frequency Rise Time Fall Time D width D width Duty Duty Delay A B Delay A B and Duty cycle Figure 4 22 shows a pulse with some of the time measurement points Figure 4 22 ...

Страница 96: ...ce B last falling edge LFR Time between Source A first falling edge and Source B last rising edge LFF Time between Source A first falling edge and Source B last falling edge Other measurements PulseCount The number of positive pulses that rise above the mid reference crossing in the waveform PulseCount The number of negative pulses that fall below the mid reference crossing in the waveform RiseEdg...

Страница 97: ...ber of samples and maximum value standard deviation of up to 8 measurement results opened last When the statistics function is turned on select Extreme Value the current value average value maximum value and minimum value will be displayed at the bottom of the screen select Difference the current value average value standard deviation and number of samples will be displayed at the bottom of the sc...

Страница 98: ... in the bottom menu the Type menu will display at the right of the screen In the right menu select Time Voltage for Type two blue dotted lines displayed along the horizontal direction of the screen two blue dotted lines displayed along the vertical direction of the screen Cursor measure window at the left bottom of the screen shows the cursor readout 4 In the bottom menu select Line Type as Time t...

Страница 99: ...tions of the vertical cursors and the waveform Move the cursors by gestures About the gestures for moving cursors see Other Operations Using Touchscreen on P23 The Cursor Measurement for FFT mode In FFT mode push the Cursor button to turn cursors on and display the cursor menu The description of the cursor menu in FFT mode is shown as the following table ...

Страница 100: ...e and frequency cursor measurement of math FFT 1 Press the Math button to display the math menu in the bottom Select FFT In the right menu select Format In the left menu turn the M knob to select amplitude unit V RMS or Decibels 2 Push Cursor to display the cursor menu 3 In the bottom menu select Window as Extension 4 Select the first menu item in the bottom menu the Type menu will display at the ...

Страница 101: ...ncy of the signals The menu is as follows Function Menu Setting Instruction Autoscale ON OFF Turn on Autoscale Turn off Autoscale Mode Follow up and adjust both vertical and horizontal settings Follow up and only adjust horizontal scale Follow up and only adjust vertical scale Wave Show Multi period waveforms Only show one or two periods If you want to measure the two channel signal you can do as ...

Страница 102: ...e oscilloscope will turn off Autoscale function To back to Autoscale Push Autoset 6 Turn off the submenu at the Autoscale menu the Autoscale is off and turn on the submenu still enters into the function 7 When video triggering the horizontal time scale is 50us If one channel is showing edge signal the other channel is showing video one the time scale refers to 50us as video one as standard 8 While...

Страница 103: ...tical Coupling Current Channel Coupling Current Vertical Scale Adjust to the proper division Bandwidth Full Horizontal Level Middle or 2 div Horizontal Sale Adjust to the proper division Trigger Type Slope or Video Trigger Source CH1 or CH2 Trigger Coupling DC Trigger Slope Current Trigger Level 3 5 of the waveform Trigger Mode Auto Display Format YT Force Stop Help Exit Pass Fail Off Inverted Off...

Страница 104: ...ted within a certain range in other words the signal can be expanded in the horizontal or vertical direction When the horizontal time base is 50ms the horizontal time base can be expanded for 4 divisions downwards Single Push this button you can set the trigger mode as single directly so when trigger occurs acquire one waveform then stop Copy You can save the waveform by just pushing the Copy pane...

Страница 105: ...e 4 In the bottom menu select Print Setup In the right menu set up print parameters The On selection of Ink Saver will print out a copy with a white background 5 Once you have connected a printer to your oscilloscope and set up print parameters you can print current screen images with a single push of the Print button on the front panel ...

Страница 106: ...elect Output in the right menu select AG Output Single channel Connect the BNC cable to the port marked Out in the back of the oscilloscope Dual channel Connect the BNC cable to the port marked Out 1or Out 2 in the back of the oscilloscope Out 1 is the output of CH1 Out 2 is the output of CH2 also can be used as the port of trigger signal output Pass Fail output Figure 5 1 Generator Output Ports D...

Страница 107: ...ttom menu as On the amplitude of the two channels can be adjusted synchronously Align Phase Select Align Phase in the bottom menu to align the initial phase of two channel signals To Set Signals 1 Push single channel or dual channel button to show the bottom menu of generator 2 Select the desired waveform in the bottom menu the corresponding menu is displayed on the right 3 The parameters can be s...

Страница 108: ...Click to increase the value of cursor position Move the cursor Click to decrease the value of cursor position To Set the Period Select Period in the right menu if Period is not displayed select Frequency and select it again to switch to Period Set the parameter in the right menu To Set the Start Phase Select StartPhase in the right menu Set the parameter in the right menu To Set the Amplitude Sele...

Страница 109: ...ters of Ramp waveform are Frequency Period Start Phase Amplitude High Level Offset Low Level Symmetry To set the Frequency Period Start Phase Amplitude High Level Offset Low Level please refer to To Output Sine Signals on page 100 To Set the Symmetry of Ramp Select Symmetry in the right menu of Ramp Set the parameter in the right menu To Output Pulse Signals The parameters of Pulse waveform are Fr...

Страница 110: ...000 1000000 1000000000 The waveform points range is 2 8192 3 Set the interpolation Select Intrpl in the right menu choose between On Off If you choose On the points will be connected with beelines otherwise the voltages between two consecutive points will not change and the waveform looks like a step up one 4 Edit the waveform points Select Edit Points in the right menu Select Point input the numb...

Страница 111: ...rowse To read a waveform stored in internal storage or USB device 1 Push single channel or dual channel button Select Arb in the bottom menu select Others in the right menu and select File Browse 2 Select the desired waveform file in internal storage FLASH or USB device USBDEVICE 3 Select Read in the right menu Built in Waveform There are 46 built in Arbitrary waveforms Steps for selecting the bui...

Страница 112: ...Stair up and stair down waveform Trapezia Trapezoid waveform RoundHalf RoundHalf wave AbsSine Absolute value of a Sine AbsSineHalf Absolute value of half a Sine SineTra Sine transverse cut SineVer Sine vertical cut NegRamp Negative ramp AttALT Gain oscillation curve AmpALT Attenuation oscillation curve CPulse Coded pulse PPulse Positive pulse NPulse Negative pulse Maths ExpRise Exponential rise fu...

Страница 113: ...re controls the built in waveform generator to sweep a sine wave across a range of frequencies while measuring the input to and output from a device under test At each frequency gain and phase are measured and plotted on a frequency response Bode chart When the frequency response analysis completes you can move a marker across the chart to see the measured gain and phase values at each frequency p...

Страница 114: ...ose knob to set waveform generator amplitude Check to enable fine tuning Note The amplitude range is 2 mVpp to 6 Vpp Chart Gain Scale Adjust scale of gain plot range from 5dB to 500dB Gain Offset Adjust offset of gain plot range from 250dB to 250dB Phase Scale Adjust scale of phase plot range from 5 to 180 Phase Offset Adjust offset of phase plot range from 180 to 180 Autoscale Autoscale gain and ...

Страница 115: ...n on the front panel to enter exit the multimeter function The button backlight will be lighted when the multimeter function is enabled Figure 6 2 Multimeter Button The bottom menu of multimeter is as below Menu Setting Description Current ACA DCA Measuring AC current Measuring DC current Voltage ACV DCV Measuring AC voltage Measuring DC voltage R C Measuring resistance Testing diode Testing for c...

Страница 116: ... the current range DMM Information Window The Multimeter Information Window is displayed on the top right of the screen 4 9 1 2 8 5 3 6 7 Figure 6 3 Multimeter Information Window Description 1 Manual Auto range indicators MANUAL refers to the measuring range in manual operation mode and AUTO refers to the measuring range in automatic operation mode 2 Measurement mode indicators A Current measureme...

Страница 117: ... measured circuit and discharge all high voltage capacitors Remove the test leads and restore the circuit to the original condition To measure a AC or DC current which is 400 mA to 10 A do the following 1 Push DMM button on the front panel Select Current in the bottom menu select it again to switch between ACA AC current or DCA DC current 2 Select Configure in the bottom menu select Current as 10A...

Страница 118: ...negative terminal cathode The cathode of a diode is indicated with a band Read the diode forward bias Testing for Continuity 1 Push DMM button on the front panel Select in the bottom menu select it until switching to 2 Connect the black test lead to the COM terminal on the back of the oscilloscope and the red test lead to the V Ω C terminal 3 Probe the test points to measure the resistance in the ...

Страница 119: ...Information Display Show hide the information window on the top right of the display 1 Select Configure in the bottom menu select Show Info in the right menu as ON The information window will be shown on the display 2 Select OFF to hide Auto or Manual Range Auto range is set as default To switch auto or manual range do the following steps 1 Select Configure in the bottom menu 2 Select Auto Range i...

Страница 120: ...rd file will be named as Multimeter_Recorder csv If a file with the same name already exists in the USB memory device it will be overwritten If you want to keep the existing file back up it to other location in advance Select STRT in the bottom menu to start recording 9 When the record time gets to the defined duration the recording will be ended If you want to end recording prematurely select STO...

Страница 121: ...t in the bottom menu to export How to chart the data You can open the CSV file using Microsoft Excel or your favorite spreadsheet application and create charts based on the data In the following steps Microsoft Excel 2010 is used as an example 1 Open Multimeter_Recorder csv in Excel Interval Measurement mode Reference value of the relative measurements 2 Select the data that you want to chart see ...

Страница 122: ...ection to choose from Using USB Port 1 Connection Use a USB data cable to connect the USB Device port in the right panel of the Oscilloscope to the USB port of a PC 2 Install the driver Run the Oscilloscope communication software on PC push F1 to open the help document Follow the steps of title I Device connection in the document to install the driver 3 Port setting of the software Run the Oscillo...

Страница 123: ... Here we set the IP address to 192 168 1 71 Figure 7 2 Set the network parameters of the computer 3 Set the network parameters of the Oscilloscope Software Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN About the IP the first three bytes is same as the IP in the step 2 the last byte should be different Here we set it to 192 168 1...

Страница 124: ...pts Reset to update the config After resetting the oscilloscope if you can get data normally in the oscilloscope software the connection is successful Figure 7 4 Set the network parameters of the oscilloscope Connect through a router 1 Connection Use a LAN cable to connect the oscilloscope with a router the LAN port of the oscilloscope is in the right side panel the computer should be connected to...

Страница 125: ...n the computer choose the Ports settings of the Communications menu item Set Connect using to LAN About the IP the first three bytes is same as the IP in the step 2 the last byte should be different Here we set it to 192 168 1 72 The range of the port value is 0 4000 but the port which under 2000 is always used so it is suggested to set it to the value above 2000 Here we set it to 3000 Figure 7 6 ...

Страница 126: ... Access Point PC should support Wi Fi 1 Set the oscilloscope In the oscilloscope push the Utility button Select Function in the bottom menu Select LAN Set in the left menu In the bottom menu set the Type item as WIFI AP and select Set in the bottom menu 2 In the right menu select SSID a keyboard will pop up You can turn and push the M knob or touch the keyboard if the LCD is touchscreen to set the...

Страница 127: ...Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN Set IP and Port to the same value in the oscilloscope in the step 4 If you can get data normally the connection is successful Figure 7 8 Set the network parameters of the Oscilloscope Software Connect with PC as Wi Fi Station The oscilloscope and PC should be on the same network LAN ...

Страница 128: ...ord In the right menu select Password use the keyboard to input the Wi Fi hotspot password 4 In the right menu select IP PORT Turn the M knob to set the PORT Here we set it to 3000 5 After setting select ON in the Status menu The value of IP is displayed in IP PORT menu Here IP is 10 1 1 126 A changing icon appears on the top right of the screen 6 Select Save set in the bottom menu to save current...

Страница 129: ... Oscilloscope Software Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN Set IP and Port to the same value in the oscilloscope If you can get data normally the connection is successful Figure 7 10 Set the network parameters of the Oscilloscope Software ...

Страница 130: ...ope is set as Wi Fi station The Android device can be connected to it through a Wi Fi hotspot Connect with APP as Wi Fi Access Point 1 In the oscilloscope push the Utility button Select Function in the bottom menu Select LAN Set in the left menu In the bottom menu set the Type item as WIFI AP and select Set in the bottom menu 2 In the right menu select SSID a keyboard will pop up You can turn and ...

Страница 131: ...n Android device enter the Wi Fi settings select the oscilloscope access point to connect enter the set password 8 In Android device launch the application 9 Click connect on the left 10 Set IP and Port to the same value as the setting in the oscilloscope Click Confirm to connect Click History to recall the history settings ...

Страница 132: ...touchscreen 3 In the right menu select Encryption push to set the Encryption type of the Wi Fi hotspot Open type needs no password WEP or WPA type need to input the password In the right menu select Password use the keyboard to input the Wi Fi hotspot password 4 In the right menu select IP PORT Turn the M knob to set the PORT 5 After setting select ON in the Status menu The value of IP is displaye...

Страница 133: ... via Wi Fi Optional 126 8 In Android device launch the application 9 Click connect on the left 10 Set IP and Port to the same value as the setting in the oscilloscope Click Confirm to connect Click History to recall the history settings ...

Страница 134: ...gain to turn CH2 on off Touch to stop continue to synchronize with the oscilloscope Time Base Horizontal Position Horizontal Position Pointer CH2 Show hide menu see below Trigger Level Share the image wave file via the installed sharing apps Save into wave file bmp png bin csv Exit APP Submit feedback Swipe from the left edge of the screen to show the trigger source menu Trigger Source Menu Trigge...

Страница 135: ...he control interface of the generator Turn on off output of selected channel Select channel Select waveform Set the parameters of current waveform Control Interface of the Generator The points will be connected with lines Adjust the brightness of grid Set the auto measurement Display Menu ...

Страница 136: ...Menu Cursor Measurement The cursor line can be dragged Check to measure voltage APP Information Enter detail setting Setting Menu Gestures Control Control the vertical position of the selected channel Show the vertical position when gesturing Control the trigger level of the trigger source ...

Страница 137: ...ation with Android Device via Wi Fi Optional 130 Control the horizontal position Pinch and spread vertically to change the voltage division of current channel Pinch and spread horizontally to change the time base ...

Страница 138: ... make the waveform optimized based on which you can further regulate the vertical and horizontal divisions till the waveform meets your requirement 2 Perform Automatic Measurement The oscilloscope can measure most of the displayed signals automatically To measure the period the frequency of the CH1 following the steps below 1 Push the Measure button to show the Measure menu 2 Select Add in the bot...

Страница 139: ...d the CH2 channel to the output end Operation Steps 1 Push the Autoset button and the oscilloscope will automatically adjust the waveforms of the two channels into the proper display state 2 Push the Measure button to show the Measure menu 3 Select Add in the bottom menu 4 In the right menu select CH1 in the Source menu item 5 In the left Type menu turn the M knob to select PK PK 6 In the right me...

Страница 140: ...out the closest trigger level and trigger type this helps user to make few small adjustments to achieve a proper trigger level and mode Here is how we achieve this The operation steps are as follows 1 Set the probe menu attenuation coefficient to 10X and that of the switch in the probe to 10X see How to Set the Probe Attenuation Coefficient on P12 2 Adjust the Vertical Scale and Horizontal Scale k...

Страница 141: ... push the Run Stop button and wait a trigger Once there is a burr occurring the instrument will trigger automatically and record the waveform during the period around the trigger time By turning the Horizontal Position knob in the horizontal control area in the panel you can change the horizontal triggering position to obtain the negative delay making an easy observation of the waveform before the...

Страница 142: ...noise level as lower as possible this would enable user to have more details about the signal The Average function offered by our Oscilloscope can help you to achieve this Here are the steps for how to enable Average function 1 Push the Acquire button to display the Acquire menu 2 Select Acqu Mode in the bottom menu 3 Select Average in the right menu turn the M knob and observe the waveform obtain...

Страница 143: ...lease operate according to the following steps 1 Set the probe menu attenuation coefficient for 10X and that of the switch in the probe for 10X see How to Set the Probe Attenuation Coefficient on P12 2 Connect the probe of channel 1 to the input of the network and that of Channel 2 to the output of the network 3 Push the Autoset button with the oscilloscope turning on the signals of the two channe...

Страница 144: ...in the I and IV quadrants that is in the range of 0 π 2 or 3π 2 2π If the principal axis of the ellipse is in the II and IV quadrants the determined phase difference angle is in the II and III quadrants that is within the range of π 2 π or π 3π 2 Example 6 Video Signal Trigger Observe the video circuit of a television apply the video trigger and obtain the stable video output signal display Video ...

Страница 145: ...he right menu 5 Select Modu in the bottom menu Select NTSC in the right menu 6 Select Sync in the bottom menu Select Field in the right menu 7 Turn the Vertical Scale Vertical Position and Horizontal Scale knobs to obtain a proper waveform display see Figure 9 7 Figure 9 7 Waveform Captured from Video Field Trigger ...

Страница 146: ...he Probe Attenuation Coefficient on P12 4 There is a waveform displayed but it is not stable Check whether the Source item in the TRIG MODE menu is in conformity with the signal channel used in the practical application Check on the trigger Type item The common signal chooses the Edge trigger mode for Type and the video signal the Video If Alternate trigger is selected both of the channel 1 and ch...

Страница 147: ...How to Implement Self calibration on P14 All specification standards can be fulfilled except one s marked with the word Typical Oscilloscope Performance Characteristics Instruction Bandwidth XDS3062A 60 MHz XDS3102 A XDS3112 100 MHz XDS3102AP 8 bits mode 12 bits mode 100 MHz 14 bits mode 25 MHz XDS3202E 200 MHz XDS3202A 8 bits mode 200 MHz 12 bits mode 150 MHz 14 bits mode 20 MHz XDS3202 200 MHz X...

Страница 148: ...chscreen should be configured Acquisition Mode Normal Peak detect Averaging Sample rate real time XDS3062A XDS3102A Dual CH 500 MS s Single CH 8 bits mode 1 GS s 12 bits mode 500 MS s XDS3102 Dual CH 500 MS s Single CH 1 GS s XDS3202E Dual CH 500 MS s Single CH 1 GS s XDS3102AP XDS3202A Dual CH 8 bits mode 1 GS s 12 bits mode 500 MS s 14 bits mode 100 MS s Single CH 8 bits mode 1 GS s 12 bits mode...

Страница 149: ...s Horizontal System Sampling rate range XDS3062A XDS3102A Dual CH 0 05 S s 500 MS s Single CH 8 bits mode 0 05 S s 1 GS s 12 bits mode 0 05 S s 500 MS s XDS3102 Dual CH 0 05 S s 500 MS s Single CH 0 05 S s 1 GS s XDS3202E Dual CH 0 05 S s 500 MS s Single CH 0 05 S s 1 GS s XDS3102AP XDS3202A Dual CH 8 bits mode 0 05 S s 1 GS s 12 bits mode 0 05 S s 500 MS s 14 bits mode 0 05 S s 100 MS s Single CH...

Страница 150: ...ading 0 4 ns Vertical system Sensitivity 1 mV div 10 V div Displacement XDS3062A XDS3102 A XDS3102AP XDS3202A 2 V 1 mV div 50 mV div 20 V 100 mV div 1 V div 200 V 2 V div 10 V div XDS3112 XDS3202E XDS3202 XDS3302 2 V 1 mV div 50 mV div 20 V 100 mV div 500 mV div 250 V 1 V div 10 V div Analog bandwidth 60 MHz 100 MHz 200 MHz 300 MHz Single bandwidth Full bandwidth Low Frequency 10 Hz at input AC co...

Страница 151: ...y Mean PK PK RMS Max Min Top Base Amplitude Overshoot Preshoot Rise Time Fall Time Pulse Width Pulse Width Duty Cycle Duty Cycle Delay A B Delay A B Cycle RMS Cursor RMS Screen Duty Phase Pulse Count FRR FRF FFR FFF LRR LRF LFR LFF Pulse Count Rise Edge Count Fall Edge Count Area and Cycle Area Waveform Math FFT FFTrms Intg Diff Sqrt User Defined Function digital filter low pass high pass band pas...

Страница 152: ... range 1 525 NTSC and 1 625 PAL SECAM Pulse trigger Trigger condition Positive pulse Negative pulse Pulse Width range 30 ns to 10 s For XDS3102AP XDS3202A the range is 8 bits mode 2 ns to 10 s 12 bits mode 4 ns to 10 s 14 bits mode 20 ns to 10 s Slope Trigger Trigger condition Positive pulse Negative pulse Time setting 30 ns to 10 s For XDS3102AP XDS3202A the range is 8 bits mode 2 ns to 10 s 12 b...

Страница 153: ...olarity Normal Inverted Trigger Condition Start Error Check Error Data Baud Rate Common Custom Data Bits 5 bit 6 bit 7 bit 8 bit I2C Trigger Trigger Condition Start Restart Stop ACK Lost Address Data Addr Data Address Bits 7 bit 8 bit 10 bit Address Range 0 to 127 0 to 255 0 to 1023 Byte Length 1 to 5 SPI Trigger Trigger Condition Timeout Timeout Value 30 ns to 10 s Data Bits 4 bit to 32 bit Data ...

Страница 154: ...formance Characteristics Instruction Full scale reading 3 digits Max 4000 count Diode 0 V 1 V Input impedance 10 MΩ On Off measurement 50 30 Ω beeping Capacitance 51 2nF 100uF 3 3 digit Voltage DCV 400mV 4V 40V 400V 1000V 1 1digit Max input DC 1000V ACV 400mV 4V 40V 400V 1 3digit 750V 1 5 3digit Frequency 40Hz 400Hz Max input AC 750V virtual value Current DCA 40mA 400mA 1 5 1 digit 4A 10A 3 3digit...

Страница 155: ... 15 W XDS3102AP XDS3112 XDS3202E XDS3202 A XDS3302 24 W Fuse 2 A T class 250 V Battery optional 3 7V 13200mAh On XDS3062A XDS3102 A the battery can last approximately 3 5 hours after charge fully On XDS3202 E XDS3302 the battery can last approximately 2 5 hours after charge fully Environment Temperature Working temperature 0 40 Storage temperature 20 60 Relative Humidity 90 Height Operating 3 000 ...

Страница 156: ...strument where the liquid crystal display will be exposed to direct sunlight for long periods of time Caution To avoid any damage to the instrument or probe do not exposed it to any sprays liquids or solvents Cleaning Inspect the instrument and probes as often as operating conditions require To clean the instrument exterior perform the following steps 1 Wipe the dust from the instrument and probe ...

Страница 157: ...x C Battery Using Guide Figure 12 1 Battery level indicator Charging the oscilloscope There will have battery power indication show on the top of panel when oscilloscope power supplied by the battery if no indication has appeared see Display on P80 indicates the battery is nearly used up Note To avoid overheat of battery during charging the ambient temperature is not allowed to exceed the permissi...

Отзывы: