ENGLISH
Owandy-Cam HD • User’s manual
20
9.2. IN CASE OF FAILURE
PROBLEMS
CAUSES
SOLUTIONS
With a computer
No image displays on the
screen and the camera
LEDs are not on.
• Defective power supply.
•connection problem.
1. Check the power supply is correctly
connected to the network and to the
connection box.
2. Check the connecting cable is correctly
connected to the handpiece and to the
connection box.
The camera switches on
but no image displays on
the screen.
•Configuration
•Driver
• Connection problem.
1. Check the camera is correctly set up in
the imaging software (please, refer to user’s
manual).
2. Check the camera is correctly detected in
the device driver (correct installation of its
driver).
3. Check the USB cable coming from the
DOCK is correctly connected to the HUB.
An image displays on the
screen, but the quality is
not satisfactory.
Camera driver configuration Check the camera configuration in the ima-
ging software (brightness, contrast, satura-
tion, etc.).
Please refer to user’s manual.
An image displays, but it is
not really clear (blurry)
• Hygienic protector.
Check the hygienic protector is correctly
positioned on the camera head.
Содержание OWANDY-CAM HD
Страница 1: ...USER S MANUAL EN NOCMEN010A 03 2021 ...