Limited monitoring duration
Description:
OsOne informs the user about the day(s)/hour(s) left before the computer will run out of disk space.
Cause(s):
The computer is almost out of free disk space.
When?
This alarm is triggered during the monitoring.
Type:
WARNING
Does it prevent from starting a monitoring?
NO
Does it interrupt the monitoring?
NO
Is it handled by OsOne?
NO
How to fix it?
By making some free disk space and/or by reducing the “Notify when free disk space is less than” value in the
“Storage” configuration screen.
How to prevent it?
•
By verifying that the remote copy and the automatic clean up options are enabled. With these 2
options enabled, experiments will be automatically cleaned up from the computer.
•
By verifying that there is no connection error preventing recorded data to be remotely copied.
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...