45
Type:
ERROR
Does it prevent from starting a monitoring?
YES
Does it interrupt the monitoring?
YES
Is it handled by OsOne?
YES
If this error
occurs during a monitoring, the “Restart tool” will:
•
Close OsOne
•
Restart OsOne
•
Restart the monitoring
If the error is not automatically resolved after 3 consecutive tries of the “Restart tool”, monitoring is stopped.
How to fix it?
•
Restart OsOne
•
Remove the BioConnect cartridge from the microscope and insert it again
•
Power off the microscope, wait a minute and power the microscope on.
If the above actions do not fix the issue, please contact Ovizio support ([email protected]).
How to prevent it?
There is no way to prevent that kind of error.
Cannot start the pump error
Description:
OsOne informs the user that there was an error when OsOne tried to start a pump sequence.
Cause(s):
OsOne tried to start a sequence while the pump was still busy
Содержание ILine F Pro
Страница 1: ...1 Ovizio Imaging Systems ILine F Pro User Manual...
Страница 12: ...12 Introduction...
Страница 16: ...16 Connect...
Страница 17: ...17 Get Started Description BioConnect drawing...
Страница 25: ...25 Press here to restart OsOne...
Страница 26: ...26 Measure...
Страница 31: ...31 Detailed views Live image Click on the Live image to enlarge it The following screen will be displayed...
Страница 39: ...39 Now you can disconnect the BioConnect pump from the Microscope pump motor...
Страница 41: ...41 Troubleshooting...
Страница 68: ...68 Configure...
Страница 76: ...76...
Страница 77: ...77...
Страница 86: ...86 Remote Access...
Страница 88: ...88 Maintenance and Servicing...