If the device has not been used for a long time or has been stored
in low-temperature environment, its pre-heat time will be signi-
ficantly longer than in normal conditions.
•
After 5 consecutive tests, allow the device to rest for 5 minutes.
•
The breathalyzer should be stored in a dry place at room tem-
perature.
•
Do not expose the breathalyzer to extreme temperature or direct
sunlight.
•
When the battery indicator blinks, the battery needs to be re-
charged.
•
If the difference between two consecutive readings is > 0.1‰
(0.01%), the test should be performed again.
•
Do NOT block the outlet during breath sampling.
•
When the “FLO” message is flashing on the display as a result of
too slow airflow or interrupted blowing, press the ON/OFF key to
restart the test. Otherwise the device switches to “ALCO” mode
after 5 seconds.
•
When you exceed the maximum time of
the test (MAX TEST), the screen shows
„500 & LOCK” and starts to blink. It means
the device needs to be calibrated. The cali-
bration is needed, so that the test results
are correct. To make the blinking „LOCK”
disappear and be able to do standard tests
again, you should have the device calibra-
ted by an authorized service.