TB
1
TB
2
TB
1
TB
2
TB
3
PLS OUT3
TEM
P
IN
TEM
P
IN
FLOW IN
0V
+
−
B
b
A
+
−
+
−
+
−
+
−
+
−
SIG
SUP
SUP
PLS OUT2
GND
POWER
N
(
−
)
H
(
+
)
PRESS IN
ANA
PLS OUT1
1/
1
OUT
MODEL
EL4001
91
25
96
20
232.4
144
136
E-886-2-E
3. INSTALLATION
3.1 Outline Dimensions
3.2 Installation
3.2.1 Installation Location
Select an installation site where:
(1) Mechanical vibration, shock and corrosive gases are
negligible.
(2) Air is dry and temperature at room temperature and
stable.
NOTE: Although the manufacturer guarantees stated
performance at ambient temperatures up to
+
50
°
C
,
it is recommended that the instrument
be placed in service at room temperature.
(3) Provide a sufficient working space behind the instru-
ment - at least 50 centimeters from the back panel of
the instrument to facilitate wiring and servicing.
3.2.2 Panel
(1) Use a rigid steel sheet with a minimum thickness of 1.6
millimeters. 3.2mm thick is recommended.
(2) If it is required to install instruments alongside each
other, dimensions in Fig.3.2 are suggested.
(3) Recommended mounting height is given in Fig. 3.3.
3.2.3 Installation
(1) Front mount the instrument through the cutout in the
panel.
(2) Fit the furnished enclosure hold-down fittings into the
top and bottom slots in the enclosure and, confirming
that the instrument is positioned horizontal, secure the
instrument to the panel with hold-down fittings (Fig. 3.1).
Optimum
Range
600
1800
450
Acceptable Range
2200
Computer
Fig. 3.1 Outline Dimensions
All dimensions in millimeters
Hold-down Fitting
Min.120
Min.220
92 ±0.8
0
+1
138
0
Fig. 3.2 Panel Cut
Fig. 3.3 Instrument Mounting Height