Vibration signals
Symbol
Vibration signal
Red
1 x
Blue
2 x
Green
3 x
Beep and vibration signals
Symbol
Vibration signal
Beep signal
Red
1 x
1 x
Blue
2 x
2 x
Green
3 x
3 x
INFORMATION
In case of a negative load > 1.5 kg or a positive load > 6 kg, the patient is informed of this fault by a vibration of
the elbow joint.
7.3.2.4 Control Options
The control programs of the elbow joint, electric wrist rotator and terminal device are configured using the respect
ive selection list (see page 56, item 4).
DynamicArm Plus control selection
Prog
Activation
Control
Version
Switching mode
1
With one linear con
trol element
The position of the product
depends on the level of the
control signal.
Simultaneous operation of the
product with a System Elec
tric Hand (via the electrodes)
is possible.
Channel 5 or 6
(green or brown
marking)
•
Flexion: pull on the linear
control element
•
Extension: reduction of
pull on the linear control
element
•
Product
stops/joint
is
locked: stop pull on the
linear
control
element
and/or release slowly
Pulling further beyond the last hold point and releasing the linear control element unlocks the elbow joint.
Prog
Activation
Control
Version
Switching mode
2
Double channel con
trol with one elec
trode
Two different control com
mands are generated for flex
ion and extension.
The speed is proportional to
the control signal.
Channel 5 or 6
(green or brown
marking)
•
Flexion: signal that begins
slowly
•
Extension: strong signal
that begins quickly
57
DynamicArm Plus 12K110N=*
Handling