4.3 General safety instructions

CAUTION

Non-observance of safety instructions

Personal injury/damage to the product due to using the product in certain situations.

Observe the safety instructions and the stated precautions in this accompanying document.

CAUTION

Signs of wear and tear on the product components

Injury due to faulty control or malfunction of the product

In the interest of preventing patient injuries and to maintain product quality, we recommend performing regu­
lar service.

Please see the service pass for more information about the service intervals. This is included in the purchased
extended warranty packages.

CAUTION

Insufficient  distance  to  HF  communication  devices  (e.g.  mobile  phones,  Bluetooth  devices,  WiFi
devices)

Injury due to unexpected behaviour of the product caused by interference with internal data communication.

Therefore, keeping a minimum distance of 30 cm from HF communication devices is recommended.

CAUTION

Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention systems, metal
detectors)

Injury due to unexpected behaviour of the product caused by interference with internal data communication.

Avoid remaining in the vicinity of visible or concealed theft prevention systems at the entrance/exit of stores,
metal detectors/body scanners for persons (e.g. in airports) or other sources of strong magnetic and electric­
al  interference  (e.g.  high-voltage  lines,  transmitters,  transformer  stations,  computer  tomographs,  magnetic
resonance tomographs, etc.).

When  walking  through  theft  prevention  systems,  body  scanners  or  metal  detectors,  watch  for  unexpected
behaviour of the product.

CAUTION

Operating the product in very close proximity to other electronic devices

Injury due to unexpected behaviour of the product caused by interference with internal data communication.

Do not operate the product in the immediate vicinity of other electronic devices.

Do not stack the product with other electronic devices during operation. 

If simultaneous operation cannot be avoided, monitor the product and verify proper use in the existing setup. 

CAUTION

Operator errors during the adjustment procedure

Injury due to faulty control of the Axon-Bus prosthesis system.

Participation in an Ottobock product training course is mandatory prior to using the product. During product
training, you will receive a password which provides you with authorisation to access the 560X500=* AxonSoft
software. To qualify for software updates, additional product training courses may be necessary.

Do not share your unlock PIN.

Use the online help which is integrated into the software.

7

AxonArm Ergo 12K501=*

Safety

Содержание AxonArm Ergo 12K501

Страница 1: ...Instructions for use qualified personnel 3 AxonArm Ergo 12K501...

Страница 2: ...1 2 2 AxonArm Ergo 12K501...

Страница 3: ...6 5 12 Adjusting the Humeral Rotation Feature 6 6 14 Installing the Pull Cable 6 7 14 Checking the Symmetry 6 8 14 Handling 7 15 Charging the Battery 7 1 15 Charging Receptacle 7 2 15 Determining the...

Страница 4: ...The movements of the Axon Bus gripping component are becoming increasingly slower 10 2 4 19 Axon 10 3 19 4 AxonArm Ergo 12K501 Table of contents...

Страница 5: ...of the elbow joint The Automatic Forearm Balance provides for harmonious movement control free swing and reduction of energy consumption It stores the energy that is released when the arm is extended...

Страница 6: ...gs 3 2 Conditions of use The product is designed for use by unilateral or bilateral amputees The product is intended exclusively for use on adults The product was developed for everyday use and must n...

Страница 7: ...ed theft prevention systems at the entrance exit of stores metal detectors body scanners for persons e g in airports or other sources of strong magnetic and electric al interference e g high voltage l...

Страница 8: ...high voltage lines transmitters transformer stations computer tomographs magnetic resonance tomo graphs etc Allow the patient to rest during the adjustment of the electrodes otherwise muscle fatigue w...

Страница 9: ...hetic glove e g 8S500 R L AxonSkin Natural when 8E500 Michelangelo hand is used Additional MyoBock components such as suction socket electrode electrode cable pull in tube and tube valve for suction s...

Страница 10: ...long co contraction on channels 1 and 2 of the AxonMaster releases or locks the lock A short co contraction switches the gripping mode of the gripper A pulse on channel 3 of the AxonMaster releases or...

Страница 11: ...Reduction in mechanical strength due to thread misalignment The circuit board and the edge of the lamination ring must be parallel to each other Left picture The EasyPlug has been snapped in properly...

Страница 12: ...isted 7 Slide the strap clamp under the lamination ring 8 Screw the strap clamp in 6 5 Setting the Preflexion The level of AFB compensation flexion assistance can be adjusted and adapted to the weight...

Страница 13: ...e forearm or the upper arm socket due to released lock The Axon Bus prosthethic system must always be flexed for donning and doffing The Axon Bus prosthethic system must be stored in flexed condition...

Страница 14: ...rntable friction is easily adjusted by turning of the external adjustment screw using a coin or similar object 6 7 Installing the Pull Cable CAUTION Removing the pull cable Injury due to malfunction o...

Страница 15: ...ing Receptacle The charging receptacle Fig 1 item 2 has the following functions Contacts for charging the battery LED display to indicate the current battery charge level LED display for providing fee...

Страница 16: ...1x short 1x long Beep Pulsating Turn the Axon Bus pros thesis solution on Push the button for one to four seconds Turn the Axon Bus pros thesis solution off Push the button for more than one second C...

Страница 17: ...that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pr...

Страница 18: ...ing General AxonArm Ergo Reference number 12K501 Dimensions mm Approx 345 mm x d55 mm to d80 mm arm extended Weight without Axon Bus terminal device g Approx 840 Forearm length as delivered mm 305 Min...

Страница 19: ...ds Probable cause The Axon Bus prosthetic system is overheated Remedy 1 Allow the Axon Bus prosthetic system to cool for a few minutes 2 Do not use the Axon Bus prosthetic system at high ambient tempe...

Страница 20: ...sion SB_2021 02 FM483 SB_210x297 Otto Bock Healthcare Products GmbH Brehmstra e 16 1110 Wien Austria T 43 1 523 37 86 F 43 1 523 22 64 info austria ottobock com www ottobock com Ottobock 647G790 0 en_...

Отзывы: