7.7 Open the Axon-Bus Gripping Component in an Emergency

This  safety  function  allows  the  Axon-Bus  gripping  component  to  be  opened  regard­
less of the control signals present. 
1)  With the Axon-Bus prosthetic system switched on, press the charging receptacle

button  and  hold  for  approximately  three  seconds  until  the  Axon-Bus  gripping
component begins to open. A pulsating beep sounds as it opens. 

2)  Releasing  the  button  immediately  stops  the  process  of  opening  the  Axon-Bus

gripping component and turns the entire Axon-Bus prosthetic system off.

7.8 Bluetooth Function

Pressing the button in the charging receptacle and holding for more than four seconds with the Axon-Bus prosthet­
ic system turned off activates the Bluetooth function of the connected AxonMaster 13E500. The blue LED lights up.
Once a connection to the PC has been successfully established, the LED flashes blue.

7.9 Beep Function

Action

Beep

1x long

Beep

2x short

Beep

3x short

Beep

1x short, 1x

long

Beep

Pulsating

Turn  the  Axon-Bus  pros­
thesis solution on

Push  the  button  for  one  to
four seconds.

Turn  the  Axon-Bus  pros­
thesis solution off

Push  the  button  for  more
than one second.

Charge  the  Axon-Bus
prosthesis solution

Connect 

the 

charging

plug, charging begins.

Stop charging

Disconnect  the  charging
plug.  Charging  ends  and
the  Axon-Bus  prosthesis
solution is switched off.

Empty battery indicator

Turn  Bluetooth  function
on

  (if  the  Axon-Bus  pros­

thesis  solution  was  previ­
ously switched off)
Push  the  button  for  more
than four seconds.

Emergency  opening  of
the  Axon-Bus  terminal
device

Push  and  hold  the  button
for  about  three  seconds
until  the  Axon-Bus  terminal
device opens.

7.10 Data transfer between the product and the PC

Product settings using the adjustment software can only be made via Bluetooth data transfer. For this purpose, a
Bluetooth wireless connection must be established between the product and the PC using the “60X5=* BionicLink
PC”  Bluetooth  adapter.  The  installation  and  use  of  the  “60X5=*  BionicLink  PC”  adapter  are  described  in  the
instructions for use included with the adapter.

16

Handling

AxonArm Ergo 12K501=*

Содержание AxonArm Ergo 12K501

Страница 1: ...Instructions for use qualified personnel 3 AxonArm Ergo 12K501...

Страница 2: ...1 2 2 AxonArm Ergo 12K501...

Страница 3: ...6 5 12 Adjusting the Humeral Rotation Feature 6 6 14 Installing the Pull Cable 6 7 14 Checking the Symmetry 6 8 14 Handling 7 15 Charging the Battery 7 1 15 Charging Receptacle 7 2 15 Determining the...

Страница 4: ...The movements of the Axon Bus gripping component are becoming increasingly slower 10 2 4 19 Axon 10 3 19 4 AxonArm Ergo 12K501 Table of contents...

Страница 5: ...of the elbow joint The Automatic Forearm Balance provides for harmonious movement control free swing and reduction of energy consumption It stores the energy that is released when the arm is extended...

Страница 6: ...gs 3 2 Conditions of use The product is designed for use by unilateral or bilateral amputees The product is intended exclusively for use on adults The product was developed for everyday use and must n...

Страница 7: ...ed theft prevention systems at the entrance exit of stores metal detectors body scanners for persons e g in airports or other sources of strong magnetic and electric al interference e g high voltage l...

Страница 8: ...high voltage lines transmitters transformer stations computer tomographs magnetic resonance tomo graphs etc Allow the patient to rest during the adjustment of the electrodes otherwise muscle fatigue w...

Страница 9: ...hetic glove e g 8S500 R L AxonSkin Natural when 8E500 Michelangelo hand is used Additional MyoBock components such as suction socket electrode electrode cable pull in tube and tube valve for suction s...

Страница 10: ...long co contraction on channels 1 and 2 of the AxonMaster releases or locks the lock A short co contraction switches the gripping mode of the gripper A pulse on channel 3 of the AxonMaster releases or...

Страница 11: ...Reduction in mechanical strength due to thread misalignment The circuit board and the edge of the lamination ring must be parallel to each other Left picture The EasyPlug has been snapped in properly...

Страница 12: ...isted 7 Slide the strap clamp under the lamination ring 8 Screw the strap clamp in 6 5 Setting the Preflexion The level of AFB compensation flexion assistance can be adjusted and adapted to the weight...

Страница 13: ...e forearm or the upper arm socket due to released lock The Axon Bus prosthethic system must always be flexed for donning and doffing The Axon Bus prosthethic system must be stored in flexed condition...

Страница 14: ...rntable friction is easily adjusted by turning of the external adjustment screw using a coin or similar object 6 7 Installing the Pull Cable CAUTION Removing the pull cable Injury due to malfunction o...

Страница 15: ...ing Receptacle The charging receptacle Fig 1 item 2 has the following functions Contacts for charging the battery LED display to indicate the current battery charge level LED display for providing fee...

Страница 16: ...1x short 1x long Beep Pulsating Turn the Axon Bus pros thesis solution on Push the button for one to four seconds Turn the Axon Bus pros thesis solution off Push the button for more than one second C...

Страница 17: ...that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pr...

Страница 18: ...ing General AxonArm Ergo Reference number 12K501 Dimensions mm Approx 345 mm x d55 mm to d80 mm arm extended Weight without Axon Bus terminal device g Approx 840 Forearm length as delivered mm 305 Min...

Страница 19: ...ds Probable cause The Axon Bus prosthetic system is overheated Remedy 1 Allow the Axon Bus prosthetic system to cool for a few minutes 2 Do not use the Axon Bus prosthetic system at high ambient tempe...

Страница 20: ...sion SB_2021 02 FM483 SB_210x297 Otto Bock Healthcare Products GmbH Brehmstra e 16 1110 Wien Austria T 43 1 523 37 86 F 43 1 523 22 64 info austria ottobock com www ottobock com Ottobock 647G790 0 en_...

Отзывы: