5.2.5 Information on Positioning and Laminating the Lamination
Anchor in the Definitive Socket
►
Note the positioning of the prosthetic components in the fully adjusted
test prosthesis and transfer to the definitive prosthesis.
5.3 Static Alignment
Static Alignment Process
+ = anterior placement / – = posterior placement (in relation to the load line)
No.
see fig. 7, see fig. 8
Required equipment:
743L100 L.A.S.A.R. Posture
To determine the load line
, position the patient on the L.A.S.A.R. Posture
as follows:
• Prosthetic foot on force plate (sufficient load:
> 35%
body weight)
• Other foot (with shoe) on height compensation plate
• The tips of the shoe and prosthetic foot are in one line
Optimise the static alignment solely by changing the plantar flexion. Adjust
ment is performed only with the distal and proximal set screws of the socket
adapter on the foot adapter.
External foot rotation:
0°
Foot flexion in m–l direction:
0°
Position of the mark on the sole relative to the load line:
0 mm
Position of the compromise pivot point (Nietert) relative to the load line:
>= -15 mm
5.4 Dynamic Trial Fitting
CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
►
Only adapt the settings to the patient gradually.
►
Explain the effects of the adjustments on the use of the prosthesis to
the patient.
During dynamic trial fitting, the alignment and settings of the prosthesis are
checked and adapted for optimum walking and running according to the
needs and abilities of the patient.
The patient has to learn the safe use of the prosthesis through intensive
training.