CAUTION
Excessive strain due to use on more than one patient
Risk of injury and loss of functionality as well as damage to the product
►
Use the product on only one patient.
►
Observe the maintenance recommendations.
NOTICE
Thermal overloading of the product
Damage due to improper thermal treatment
►
Do not carry out any heat treatment on aluminium at temperatures
above
150 °C
.
►
Do not carry out any heat treatment on steel at temperatures above
300 °C
.
►
Prior to thermal treatment, remove all temperature-critical components
(such as plastic parts).
NOTICE
Exposure of the product to unsuitable environmental conditions
Damage, brittleness or destruction due to improper handling
►
Avoid storage in condensing ambient humidity.
►
Avoid contact with abrasive substances (e.g. sand, dust).
►
Do not expose the product to temperatures below -10 °C (14 °F) or
above +60 °C (140 °F) (e.g. sauna, excessive sunlight, drying on a
radiator).
5 Preparing product for use
Orthosis alignment
►
For the upper and lower leg area, select suitable materials to reduce the
torsional stress on the orthosis.
►
Record the parts and lot number in the patient file to enable tracing in
accordance with product liability law.
►
Use the 743Y57=1-6 alignment insert for the correct alignment of the
orthotic joints.
►
The orthotic joints are produced with a standard 7° knee flexion angle.
When adjusting orthotic joints, do not exceed a knee flexion angle of 10°
for aluminium and 15° for steel.
►
When using thermoplastic material, additional reinforcement is required
in the area of the thigh and calf.