Place the spring bolt on the module and tighten it with a
Torx wrench until the spring dummy fits in the module
with no play.
3.9 Reaction-module – static alignment
Put the orthosis on the patient.
Use the knee pivot point gauge to determine and mark
the compromise pivot point according to Nietert on the
knee joint.
24
Procedure
17AD100