MIDIMATE 602 & 622 Diagnostic Audiometers
26
The display shows
Printing ...
during printout. In MIDIMATE 622, versions 3.0
and 3.1, the
Ext. Range HL
lamp flashes during printout.
8.2.3
Viewing stored test results
To view the stored thresholds, do as follows:
1.
If required, reselect the test mode with the same function and output as the
thresholds to be viewed by pressing
Menu
and then using the
Select Func-
tion
sequence.
2.
Press
Softkey 3
,
View
.
3.
If required, toggle
L/R Shift
to select the appropriate ear.
4.
Turn the
Frequency
or
Hearing Level
knob to change the frequency of the dis-
played threshold.
8.2.4
Printing separate Air and Bone Conduction audiograms
MIDIMATE 622 prints out both Air and Bone Conduction thresholds on the same
audiogram (if there is one). To obtain separate audiograms, do as follows:
1.
Test the patient with the first transducer type to be used.
2.
Print out the stored thresholds that were registered.
3.
Clear the stored data by pressing
Softkey 5
,
Clr-All
, in the
Printer Menu
.
4.
Test the patient with the second transducer type.
5.
Print out the stored thresholds that were registered with the second transducer
type.
With
Softkey 1
,
Clr-L
, and
Softkey 2
,
Clr-R
, stored left and right thresholds can be
retained or cleared independently of each other.
8.3
Setting up printout
To set up printout on the MIDIMATE 622, do as follows:
1.
Switch off the MIDIMATE 622.
2.
Press
Softkey 3
while you switch on the power.
3.
Release
Softkey 3
when the
User Progr. Parameters
menu is displayed.
4.
Press
Softkey 5
,
Printer
, to display the
Printer Options
. This menu gives
access to five sub-menus.
5.
Press
Softkey 1
,
Select
, to display the
Select Printer
menu.