MIDIMATE 602 & 622 Diagnostic Audiometers
24
7
Programming a User Test
7.1
Accessing user test programming mode
To access User Test programming, do as follows:
1.
Switch off the MIDIMATE 602/622.
2.
Press
Softkey 3
while you switch on the power until the
User Progr. Pa-
rameters
menu is displayed.
3.
Press
Softkey 3
,
Progr.
to select the
Enable User Programming
menu.
4.
Press
Softkey 1
,
Yes
, or
Softkey 2
,
No
, as required.
5.
Press
Menu
to quit the
User Progr. Parameters
menu and execute the
normal start-up procedure ending in User Test 1.
7.2
Programming a user test
A user test will store choices such as
Output
and
Stimulus Side
, the test fre-
quency and the hearing and masking levels, and also options such as
HLL
and
PTS
.
To program or change a user test, do as follows:
1.
Activate user test programming as described in §7.1.
2.
Press
Menu
once or twice to quit User Test 1 and display the
Select Function
menu or select an existing User Test to be changed.
3.
Set up or change the selected User Test as for manual setup by menu (§3.2).
4.
To check that user test programming actually is activated, or to save the current
user test, press
Softkey 3
to display the
Options Menu
. This menu should show
Progr.
above
Softkey 3
.
5.
Wait 5 sec or press
Menu
to return to the test mode, or press
Softkey 3
,
Progr.
,
to display the
Programming User Test!
menu.
6.
Press
Softkey 5
,
Quit
, to change or check the test setup, or
Softkey 1
,
Con-
tinue
, to assign a number to the User Test.
7.
Press
Softkey 1
,
2
,
3
,
4
or
5
to save the user test as
Test1
,
Test2
,
Test3
,
Test4
or
Test5
respectively, and exit from user test programming to test mode with
the setup of the user test you have just stored.