Water flow in the stimulation mode (low flow) 150 ml/min: standard unit 150/450 ml/min (adjustable)
250 ml/min: variant 250/450 ml/min (adjustable)
400 ml/min: variant 400/400 ml/min (adjustable)
500 ml/min: variant 500/500 ml/min (adjustable)
Operating pressure
at least: +2000 hPa (2 bar); max: +6000 hPa (6 bar)
Connections
Power connection; control output for a nystagmograph; equipotential
equalization; water supply G 3/4" external thread; water drainage G
1/4" external thread; connection for double-barreled hose
Operating time
Continuous operation
Fuses
quick-acting 8 A (220-240 V~); quick-acting 15 A (110-127 V~)
Protective earth conductor resistance
0.1 Ω max.
Earth leakage current
0.5 mA max.
Enclosure leakage current
0.1 mA max.
Patient leakage current
0.1 mA max.
Ambient conditions
Transport/storage
Operation
-20...+50° C; 5...90% humidity without condensing, air pressure
700...1060 hPa
+10...+35° C; 20...80 % humidity without condensing, air pressure
700...1060 hPa
Dimensions HxWxD
14.5 x 37 x 31.5 cm
Weight
approx. 5.5 kg
Protection class (EN 60601-1)
I
Degree of protection
Type B
Protection category
IPX0
Classification acc. to Annex IX EEC directions
93/42/EEC
IIa
CE marking
CE 0124
Rules applied
EN 60601-1: 1990 + A1:1993 + A2:1995
EN 60601-1-2: 1993 (EMV / EMC)
Canadian Classification Device group
Ear, nose, throat
PNC
77ETP
Risk Class
2
Description
STIMULATOR, CALORIC (WATER)
Issue of the Technical Specifications: 01.02.06
28
Otometrics - Aquamatic II
9 Specifications