PRODUCT SETTINGS AND CONFIGURATION
OI-6000K-X-X-X-NXP-X OPS_GUIDE_REV 1.0
28
3.7
FAULT TERMINAL FAIL-SAFE SETTING
The Fault Terminal Fail-Safe status behavior can be adjusted to either activate during a fault condition or
deactivate during a fault condition. The default setting is to deactivate during a fault condition, this setting
should only be adjusted if the opposite behavior is desired.
3.7.1
FAULT TERMINAL: FAIL-SAFE SETTING
1.
Use the
ADD
and
SUB
buttons to select the desired fail-safe setting for the Fault Terminal.
Select “Yes” to turn the fail
-
safe setting on, or select “No” to turn the fail
-safe setting off.
2.
Press the
MENU
button to select the desired setting and to advance to the Calibration Method
setting screen.
3.8
CALIBRATION METHOD
The calibration method selection allows you to choose how you calibrate the sensor element. Manual
calibration is the default method for all gas types.
◼
Manual calibration lets you use the
ADD
and
SUB
buttons during calibration to match the reading
shown on the screen to the value of the gas being applied.
◼
Auto calibration will set the reading, after a predetermined amount of time, during calibration to the
value entered during the auto calibration setup process.
1.
Use the
ADD
button to select manual calibration and the
SUB
button to select auto calibration.
2.
Press the
MENU
button to advance to the Modbus Address Setting screen.