Day to Day Usage

Prior to daily use, the detector will prompt the user to perform a self test.  This process is a simple and effective way to 
ensure safe operation of the detector. During the “Self Test” the audio, visual, and vibrating alarms are activated and the 
sensor is tested. Below is a step by step process for performing the test:

Screen Display

Detail

Screen 1:

When the “Test” Icon appears in the upper left hand corner, 
a self test is required. Press the button on the front of the 
detector to perform the test. 

Screen 2:

After pressing the button, the following screen will appear. 
During the self test, ensure that the following occurs: The 
unit emits one audible beep; all LEDs light up; the unit 
vibrates; all LCD display elements appear.
Screen 3:

After the full element LCD screen, the low alarm and high 
alarm set points will be displayed.  NOTE:  These alarm set 
points can be adjusted using the IR Link or Calibration 
Station Dock configuration options.

Factory Standard Alarm Set Points are:
H2S = Low 10 ppm / High 15 ppm
CO = Low 35 ppm / High 200 ppm
O2 = Min 19.5% / Max 23.5%
Screen 7: (see steps 4-6 for additional screen displays)

When a self test is successful, the unit will turn to the 
original screen, displaying a check mark in place of where 
the test icon was previously displayed, and one audible 
beep will sound.

The unit will by default prompt another self test in 20 hours 
from when the button was pressed.

  

   NOTE:  This value can be changed via the IR Link      

     software  from 8-20 hours.  See the IR Link Quick 
     Reference Guide for details.

6

Содержание OI-2100 TOCSIN

Страница 1: ...Monitor ________________________________________ Operation Manual Revision 2 1w _______________________________________________________________________________________________ ________________________...

Страница 2: ...tisseement 4 First Time Activation 5 Display Details 5 Day to Day Usage 6 Additional Steps 7 Failures FAQ s 8 Alarms 9 Alarm Types 9 Alarm Set Points 9 Event Log 10 Bump Test Interval bUP Clearing the...

Страница 3: ...el OI 2100 TOCSIN Single Gas Personal Monitor This document should be read before initial operation of the product Should a question arise during the use of the product this document will serve as a f...

Страница 4: ...ny obstructions i e debris or blockage Perform the Self Test to ensure proper operation of visual audible and vibrating alarms Confirm receipt of a check mark in the manual test is to be performed mak...

Страница 5: ...TE The display mode can be changed in the IR Link software with the Sensor Reading and Life Remaining user options WARNING Users must familiarize themselves with the icons in both non alarm and alarm...

Страница 6: ...beep all LEDs light up the unit vibrates all LCD display elements appear Screen 3 After the full element LCD screen the low alarm and high alarm set points will be displayed NOTE These alarm set point...

Страница 7: ...software Screen 5 If the detector has been exposed to gas exceeding the low alarm set point a value will appear with max next to it This represents the peak value highest that the detector has seen Af...

Страница 8: ...the LCD will display EO5 In case of an EO5 screen discontinue use and contact Otis Instruments for a replacement detector Along with the battery the sensor is continuously monitored during normal oper...

Страница 9: ...hen the unit has eight hours remaining it will begin to beep flash and vibrate intermittently NOTE To end the alarms press the button down Once the unit has reached the end of its operating life the d...

Страница 10: ...re from 1 to 365 days NOTE The unit default is to have no bump interval programmed If a detector is due for a bump test the unit display will alternate between the months remaining and bUP In addition...

Страница 11: ...ay Detail O2 CAL NOTE Only perform O2 calibration in normal ail 20 9 Oxygen that is free of hazardous gases 1 Press and hold the yellow button for four seconds 2 The display screen will show CAL and t...

Страница 12: ...dible minimum 95dB Battery Life 24 months of operation 2 minutes of alarm per day Event Log Storage Last 25 events Newer events replace older events Certifications Class I Groups A B C D T4 ETL 400319...

Страница 13: ...Proper storage installation calibration use maintenance and compliance with the user s manual instructions and any other applicable recommendations of Otis Instruments 2 The buyer promptly notifying...

Страница 14: ...Otis Instruments Inc 2200 E Villa Maria Dr Bryan TX 77802 Service Department 903 566 1300 Corporate Office 979 776 7700 service otisinstruments com www otisinstruments com 14...

Отзывы: