64
OticOn ActO
FITTING GUIDE
65
Buttons and Indicators
In the Button and Indicators you define the functionality of
the available buttons, volume control and beeps on the
instruments.
This is the place where you choose whether you want the
buttons to affect both instruments at the same time by clicking
the Binaural Coordination button next to the operation you want
handled.
You have the option of including a Mute function in the buttons.
By default this is set to off. When turned on, the Mute function
can be activated by the client with a very long push on
the button.
Double Button on BTE and RITE
This switch can move the program cycle forward or backward or
the volume control louder and softer for each instrument.
Genie gives you the flexibility to program the switch for program
changes, volume changes or both. You can also decide with the
client if the switch control should be programmed the same to
both hearing instruments. Some clients prefer to have one ear
be for volume while the other is for program changes. What ever
option is chosen, “binaural coordination” must be enabled in
Genie to ensure the changes are applied to both instruments.
Instruct your clients to use a short push for program changes
or volume changes. If the instrument has been programmed
for both functionalities instruct the client to use a short push
for volume changes and a medium long push for program
changes.
In the
User instruction
task you can print the button configu-
ration and include it into the Instructions For Use to provide
the information for your client.
Beeps
The Beeps tab is used to define the audible indication for
program and volume control changes as well as decide when
there should be audible indication in the instruments.
Note: Remember to fill out the programmed button controls in
the Instructions For Use before handing it out to the client.