Component maintenance
Maintenance of the humidification cup:

 In order to ensure the cleanness of the humidification 

cup, only can sue pure water or cold boiled water, and should be replaced every day. The 
humidification cup is recommended to be cleaned once a week.

Maintenance of oxygen pipe and oxygen inhaler:

 It is recommended to clean once a week, 

and the use of the oxygen tube should not exceed half a year.

Maintenance of air filter cotton: 

The main function of air filter cotton is to filter and absorb dust 

and impurities in the air. After a long time, it will form an air intake blockage and affect the 
oxygen production effect of the oxygen concentrator: when the machine works continuously for 
200 hours, “Clean Filter” will be  displayed on the screen, please replace the filter in time. The 
words “Clean Filter” will continuously display for 1 hour.      

Working principle

 Air

Primary filter

Secondary filter

Tertiary filter

compressor

Earhook oxygen inhaler

Molecular integrator

Humidification

cup

Bacterial filter

Troubleshooting

Number

Common

troubleshooting

The oxygen concentrator 

is not working properly 

and the power indicator 

is not on.

The screen display is 

normal, the oxygen 

machine does not 

emit a normal start 

sound

Display normally and 

can hear the sound 

of normal working.

possible reason

The power cord is not connected to 

the power socket

Check if the plug on the power 

socket is connected properly.

The power cord is not connected to 

the oxygen concentrator or the 

power cord is broken.

Oxygen concentrator overheat 

protection

Oxygen concentrator computer 

board crash

Compressor malfunction

Check the bottom strap

Inlet window filter cotton is too dirty, 

blocked

Output hose bent and blocked

Humidification cup is not tight

Internal failure of the oxygen 

concentrator

The silencer is not installed correctly

Unplug and reinsert the power 

socket or replace a power cord

Turn off, restart after 5 minutes.

 Cut the strap and remove it

Reinstall correctly after disassembly

Remove and re-install

Cut off the power switch, Replacement  

filter cotton

Contact the distributor or 

manufacturer of the home oxygen 

concentrator

Cut off the power switch, Replace 

the output hose or unfold the 

curved hose

Contact the distributor or manufacturer 

of the home oxygen concentrator

Make sure that the oxygen 

concentrator is placed away from 

curtains, heaters, heaters and 

fireplaces. And confirm that there is 

at least 20 cm clearance between 

all sides of the machine and the 

wall or other objects.

Power socket is out of power

Replace the socket with power

Solution

Loud noise

Note

If The above methods can’t solve the faultiest. Please contact the supplier or 

manufacturer. No one can access or change the internal structure and program of the 
machine without the manufacturer’s authorization.

Page 11 of 13

Page 12 of 13

Содержание SYK-608-2

Страница 1: ......

Страница 2: ...t surface such as a bed or a sofa that will cause tilting or sinking Prevent shut off or oxygen concentration decline from over temperature caused by blockage of inlet and outlet Fire warning or explo...

Страница 3: ...e body improve brain vitality and cardiopulmonary function and improve the quality of sleep Normal Operating condition Ambient temperature range 0 40 Relative humidity range 80 Atmospheric pressure ra...

Страница 4: ...afety ground socket that has a power output Do not use the extended cord Do not bend the oxygen tube when the machine is working it will cause the oxygen tube to fall off If this happens please re ins...

Страница 5: ...ontinuous running state of the machine 6 Anion Touch the Anion key to turn on and off the negative ion work 7 Smart Alarm Touch the Smart Alarm button and broadcast the assistance Seeking information...

Страница 6: ...ization function 1 The atomization tube is inserted into the bottom of the atomizing cup 2 Put the cup cover and unscrew the spray to spray medicine Note liquid limit 3 note that the liquid not excess...

Страница 7: ...r cord is not connected to the power socket Check if the plug on the power socket is connected properly The power cord is not connected to the oxygen concentrator or the power cord is broken Oxygen co...

Страница 8: ...uct exceeds warranty period the company will charge appropriate cost according to the replacement parts 2 Maintenance should be with a receipt and warranty card in order to determine the warranty peri...

Отзывы: