background image

10 of 20

Fig. 9:  wristband battery housing

Batteries are fitted into the wristband battery housing and 
secured by the holding catch at the wrist side of the wrist-
band. Release the holding catch to remove the battery.

The batteries provided are designed to work 
specifically with the Touch Bionics i-digits™ 
wristband and are not to be used with any 
other device.

Only Touch Bionics batteries are approved 
for use with i-digits™, use of alternative bat-
teries will invalidate the warranty.

Charging time from full discharge is approximately 2 hours. 

i-limb

®

 Power Pack batteries for i-digits™ should only be 

charged using the Touch Bionics battery charger (fig.7) 
supplied. Place the batteries in the charger as illustrated. 
Insert the charger lead from the mains charger into the 
charger base unit. Insert the mains charger into the power 
outlet (fig.10).

On the base plate of the charger base unit you will see the 
various light sequences:

•  Middle light on: Charger is plugged in
•  2nd and 5th lights blinking Green: Batteries are 

charging

•  2nd and 5th lights solid Green: Batteries are charged
•  1st and 4th Red lights on: Battery fault, unplug and 

try again. If lights continue to illuminate, contact your 
clinical team.

Fig. 10: charger base unit

Do not pull cable to remove the lead.

Switch the device OFF to preserve battery 
power when not in use. 

Replace the battery annually for optimal 
performance. 

To ensure i-digits is continually functional, 
charge at the end of each day.

Содержание i-digits

Страница 1: ...i digits User Manual...

Страница 2: ...ary to ensure you have the latest copy of this document For further training information and videos demonstrating use of i digits devices in a variety of functional activities visit http training touc...

Страница 3: ...8 2 3 Battery Charging 9 3 0 Covers 11 3 1 Cover Options 11 3 2 General Precautions 11 4 0 Grip Review 12 4 1 Grip Options 12 5 0 Support Information 14 5 1 Storage and Maintenance 14 5 2 Troubleshoo...

Страница 4: ...o individuals with any 3 4 or 5 digit loss while patients or users with 1 or 2 digit loss are also indicated when the digits of loss are either the thumb or the main digits of opposition namely the in...

Страница 5: ...ics Customer Support team to resolve the matter as quickly as possible To best protect the hand when carrying objects the weight should be distributed evenly across the digits as close to the knuckles...

Страница 6: ...other grip icon Speed boost not available on i digits access is adjusted using a simple slider bar on the my i limb app This increased speed can provide benefits of a more responsive action more natur...

Страница 7: ...0 Control and Batteries Fig 3 Prosthesis overview Motorized Digit Knuckle Wristband and batteries This illustration above depicts an i digits for users with sound thumb and 4 i digits Palmer view Dor...

Страница 8: ...You may experience some minor redness on your residual hand but this should not last very long after you remove the device If the device is causing you discomfort or affecting you using your i digits...

Страница 9: ...low on charge If the battery is low on charge the LED will illuminate RED The LED will remain illuminated until the battery is adequately charged Only use supplied Touch Bionics charger to charge batt...

Страница 10: ...onics battery charger fig 7 supplied Place the batteries in the charger as illustrated Insert the charger lead from the mains charger into the charger base unit Insert the mains charger into the power...

Страница 11: ...lation Covers will be provided for i digits device Digit covers will be ready to fit and are designed to cover each digit Touch Bionics approved digit covers must be used with each digit of the devic...

Страница 12: ...ide grip Thumb Precision Pinch Opened middle ring and little finger remain fully opened and switch off Thumb automati cally moves to a partially closed position Index finger will move to provide grip...

Страница 13: ...er defined position The user can choose to keep certain digits active and switch oth ers off Grasp fingers adopt a pre flexed position All fingers function in squeeze mode for rapid grasp and release...

Страница 14: ...formation 5 1 Storage and Maintenance Always turn off your i digits device when not in use Aim to charge the battery each day after use Replace the batteries every 12 months Ensure your i digits is se...

Страница 15: ...d contact One digit does not operate Check if the digit operates correctly using the Hand Health Check in my i limb app Device stops halfway during an action Contact your Prosthetist Prosthesis opens...

Страница 16: ...ver Damaged covers must be replaced or repaired by a quali fied Touch Bionics technician or technical partner Only approved Touch Bionics accessories and tooling may be used with i digits Do not use a...

Страница 17: ...be returned to customer service Please forward to Touch Bionics as per the contact information on the back page of this manual It is recommended that the above information is also included in the user...

Страница 18: ...ronments listed below The user of i digits device should assure that it is not used in such environments Condition Level Maximum temperature 70 c Minimum temperature 40 c Hazardous Area Classification...

Страница 19: ...nts and or batteries that should not be disposed of in regular waste at the end of its usable life To prevent possible harm to the environ ment or human health from uncontrolled waste disposal users a...

Страница 20: ...751 03 Uppsala Sweden Tel 46 1818 2200 Fax 46 1818 2218 info ossur com ssur Iberia S L U Calle Cal ndula 93 Miniparc III Edificio E Despacho M18 28109 El Soto de la Moraleja Alcobendas Madrid Espa a T...

Отзывы: