Disinfection, Cleaning, Sterilization
Sterilization and storage
W&H recommends sterilization according to EN 13060, class B
– Note the instructions of the unit manufacturer.
– Clean before sterilizing.
– Wrap motor and accessories in sterile goods packing according to EN 868-5.
– Make sure that you only remove dry sterile goods.
– Store sterile goods dust-free and dry.
For sterilization with a drying program: Remove the motor protective cover.
For sterilization without a drying program: Use the motor protective cover (see Page 26).
Approved sterilization procedures
Follow your country-specific directives, standards and guidelines.
Steam vacuum sterilization according to EN 13060, using a sterilization holding time of a minimum of
3 minutes at 134 (+ 3) °C (273.2 + 5.4 °F)
or
Steam gravity sterilization using a sterilization holding time of a minimum of 3 minutes at 134 (+ 3) °C (273.2 + 5.4 °F)
Before starting operation again
Wait until the motor and cable are completely dry.
Moisture in the plug or motor can lead to a malfunction! (Risk of short circuit)