Disinfection, Cleaning, Sterilization
Motor with cable
Do not twist and kink the motor cable! Do not coil it too tightly!
This instruction is not necessary when a sterilization cassette
is utilized (optional accessory).
Pre-disinfection
– If heavily soiled: Clean first with disinfectant cloths.
Manual cleaning
– Rinse under demineralized water (< 38 °C) with the aid of a brush (not recommended in U.K.)
– Then remove any liquid residues (absorbent cloth, blow dry with compressed air)
Disinfection
– Disinfection with surface disinfectants,
wiping
disinfection is recommended.
–
Use only surface disinfectants which do not contain chlorine and which are certified by officially recognized institutes
.
For USA:
Use EPA registered surface disinfectants.
– Note the dosage and the action time of the
surface disinfectants used
!
Before disinfection and cleaning
push the motor protective cover
onto the motor attachment!