9
IV.
The UV lamp is a mercury vapour lamp. The user may be exposed to mercury if
the lamp glass breaks. If this happens, please abide by the following rules to
minimise exposure:
•
Ensure the power is disconnected to avoid the risk of electrocution.
•
Store the container and UV Ozone Cleaner in a fume hood and leave the
room under ventilation for at least 15 minutes.
•
Using cut-resistant gloves, collect any stray pieces of glass in a sealed
container. Dispose of the glass at a suitable collection point.
V.
The UV Ozone Cleaner has a high voltage source. Do not attempt to open the
unit.
3.5
Power Cord Safety
I.
Emergency power disconnect options: Use the power cord as a disconnecting
method and remove it from the power source. To facilitate disconnect, make
sure the power outlet for this cord is readily accessible to the operator.
II.
Only use the power cord (and transformer if using 110 V mains supply) supplied
with the UV Ozone Cleaner. Using an unearthed plug may result in serious injury
or death.
3.6
Servicing
If servicing is required, please return the unit to Ossila Ltd The warranty will be invalidated if:
•
Modification or service has taken place by anyone other than an Ossila engineer.
•
The Unit has been subjected to chemical damage through improper use.
•
The Unit has been operated outside the usage parameters stated in the user
documentation associated with the Unit.
•
The Unit has been rendered inoperable through accident, misuse, contamination,
improper maintenance, modification, or other external causes.
3.7
Health and Safety
–
Installation
I.
High-intensity, mercury vapour lamps can generate extreme heat and
temperatures. Furthermore, ozone is a powerful oxidising agent and can react
explosively with combustible materials. Keep flammable materials a minimum
of three feet away from operating equipment.