5.13 Connect the Water / Waste Hoses
Humidifier Water Fill
The mobile unit contains a water storage tank for the humidifier. This tank is located in the
equipment room and must always contain water to insure the specified humidity level
remains constant. There are two options for filling the tank.
•
A ¾” G.H.T. male threaded hose connection is located under the front kick of the
mobile unit, below the air conditioning units.
•
A fill port is located on the water tank itself for manual fill capability.
1. Remove the cap that covers the connection.
2. Attach the supplied water hose to this connection.
3. Attach the other end of the hose to facility provided faucet.
4. Turn on the water at the faucet.
5. This will fill the water tank for the mobile unit.
6. In the event that the water tank is overfilled, an overflow drain has been
provided that exits outside of the mobile unit.
Fresh Water Supply Requirements (Optional)
The mobile unit will be supplied with a ¾” diameter, 20’-0” long hose terminated with a ¾”
G.H.T. male threaded hose connector located on the left side of the mobile unit. Refer to
Figure 20: Fresh Water Connection
. The facility must provide a ¾” female connecter and
a water supply that meets the following specifications:
•
A flow rate of 5 gallons per minute.
•
45-60
PSI.
•
A maximum temperature of 70°F.
Waste Water Connections (sink option)
A 1½” IPS male connection is required to sink wastewater.
Any pipes and drains within 20’-0” of the magnet isocenter should be of non-ferrous
materials such as PVC, copper, or brass. All plumbing must comply with all applicable
codes.
The mobile unit is supplied with a 20’-0” long 1” diameter hose that is terminated with a 1”
male threaded connector for sanitary wastewater drainage located on the left side of the
mobile unit. The facility must provide means of sanitary wastewater drainage from the
system, which complies with all local applicable codes.