oserio BHG-208 Скачать руководство пользователя страница 15

15 

 

frequency range

b

Interference may occur in the 

vicinity of equipment marked 

with the following symbol:  

 

NOTE1   At 80 MHz and 800 MHz, the higher frequency range applies.  

NOTE2  These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.  

a    Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 

environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the product  is used 

exceeds the applicable RF compliance level above, the product should be observed to verify 

normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as re-orienting or relocating the product.  

b    Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.   

 

Recommended separation distance between portable and mobile RF communications 

equipment and the BHG-208 Cardio Scale 
The product is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The customer or the user of the product can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile 

RF communications equipment (transmitters) and the product as recommended below, 

according to the maximum output power of the communications equipment.  

Rated maximum output 

power of transmitter   

 

W   

Separation distance according to frequency of transmitter 

m   

150 kHz to 80 MHz   

d =1,2√ 

80 MHz to 800 

MHz   

d =1,2√ 

800 MHz to 2,5 GHz   

d =2,3√ 

0,01 

  0,12 

  0,12 

  0,23   

0,1  

0,38 

  0,38 

  0,73   

1   

1,2  

1,2  

2,3  

10  

3,8  

3,8  

7,3  

100  

12 

  12 

  23  

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance d in metres (m) can be estimated using the equation applicable to the 

frequency of the transmitter, where p is the maximum output power rating of the transmitter in 

watts (W) according to the transmitter manufacturer.  

NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.   

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 

by absorption and reflection from structures, objects and people.  

Содержание BHG-208

Страница 1: ...1 User Manual BHG 208 Cardio Scale Please keep the instruction manual at hand all the time for future reference...

Страница 2: ...is user manual is protected by international copyright law All content is licensed and usage is subject to written authorization from Charder Electronic Co Ltd hereinafter Charder Charder is not liabl...

Страница 3: ...escription 6 A Overview 6 B Inserting Batteries 7 III How to Use 8 A Self calibration 8 B Conducting measurement 8 IV Caring for your product 10 A Use Storage 10 B Service and maintenance 10 V Trouble...

Страница 4: ...le flat solid non slippery surface Place the product in an area free from direct sunlight heating equipment high humidity or extreme temperature change Never submerge in water Usage on soft surfaces e...

Страница 5: ...ice with operational Bluetooth and Wi Fi is required to download the o Care app and retrieve measurement results from the product QR Code QR Code Google Play Android Devices App Store Apple Devices Af...

Страница 6: ...6 II Product Description A Overview 1 LED screen 2 Measurement platform 3 Feet 4 Battery cover...

Страница 7: ...over Insert 3 AAA batteries Replace battery cover NOTE Ensure batteries are inserted in correct orientation If batteries are inserted incorrectly product will not function If product will not be used...

Страница 8: ...no carpet To ensure accuracy we recommend performing self calibration if product is moved Open o care app on smartphone and log in Results will be automatically transferred after measurement so it is...

Страница 9: ...results 1 If o care app is not open on smartphone results will not be transferred Product will automatically shut down 30 seconds after completion of heart rate measurement 2 Ensure that Bluetooth is...

Страница 10: ...o not use detergents or cleaners to clean the product This may damage the coating protecting the electrodes used for body composition measurement Wipe the platform and display using clean soft cloth A...

Страница 11: ...s product s maximum capacity Stay still and avoid movement during measurement Data transfer error Data sent from smartphone to product resulted in error Try measurement again Data transfer error Data...

Страница 12: ...Product 290 W x 300 D x 25 H mm Screen 76 W x 42 5 D mm Operation Environment Temperature 5 35 Relative Humidity 30 85 Storage Environment Temperature 20 60 Relative Humidity 10 95 Capacity Graduatio...

Страница 13: ...ations flicker emissions IEC 61000 3 3 Compliance Guidance and manufacturer s declaration electromagnetic immunity The BHG 208 Cardio Scale is intended for use in the electromagnetic environment speci...

Страница 14: ...the electromagnetic environment specified below The customer or the user of the product should assure that is used in such an environment Immunity test IEC 60601 test level Compliance level Electroma...

Страница 15: ...rtable and mobile RF communications equipment and the BHG 208 Cardio Scale The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The custom...

Страница 16: ...sion Yangzhou Co Ltd No 28 Xinyang Rd Development Zone Yangzhou City Jiangsu Province 225009 P R China on behalf of Charder Electronic Co Ltd No 103 Guozhong Rd Dali Dist Taichung City 412 Taiwan R O...