Caution
•
Use only the cleaning agents prescribed for cleaning the device.
See
Recommended cleaning agents
►
19
.
Frequency
We recommend that you set up a schedule for cleaning the device and accessories such as probes and/or earphones.
Prerequisites
•
Before cleaning, switch off T830/T840 and disconnect it from any external power source.
•
If needed, unplug the contralateral phone from the device.
Cleaning the probe tip
See
Cleaning the probe and probe tip
►
19
.
Disposal
There are no special requirements for the disposal of disposable articles such as eartips, i.e. they can be discarded accord-
ing to local regulations.
10.1.1
Recommended cleaning agents
Caution
•
Use only the cleaning agents prescribed for cleaning the device.
For cleaning the device, we recommend that you only use non-alcohol-based disinfectant wipes (e.g. Audio wipe) or a
cloth dampened lightly with a recommended cleaning agent to ensure proper infection control and maximum lifetime of
the device.
The following chemical solutions are recommended:
Cabinet surface and probe
•
Non-alcohol-based disinfectant wipes (e.g. Audio wipe)
•
Lukewarm water with a mild detergent
Caution
•
If plastic parts are soaked in a cleaning agent they will deteriorate.
10.1.2
Cleaning the probe and probe tip
Although the probes are designed to be easily cleaned, care should be taken to make sure that they last a long time.
Note
•
Check the sound channels in the probe tip every time you have used the probe. Even small amounts of cerumen
or vernix can block the sound channels. Clean the sound channels if needed.
Note
•
Accurate testing is only guaranteed if you use the eartips approved specifically for the device.
19
User Guide
Inmedico A/S - T830/T840