Changing the hearing level offsets: 
The hearing level offset values are stored in the audiometer’s memory. They can be changed if desired, for 

example to use another standard, or if the ISO standard is changed. This is done in a special mode activated 
by  holding  down  the  keys  ‘FREQUENCY  DOWN’,  ‘MASKING  ATTENUATOR  DOWN’  and 

‘MASKING ATTENUATOR UP’ while powering on the device. When in this mode, the frequency keys 
select  the  frequency  to  be  changed,  and  the  tone  attenuator  does  the  actual  change.  The  offset  value  is 

displayed in the Identification display. 

 
 

 