General information: 
Due to tolerances in electronics and transducers (telephones and bone conductor), it is necessary to calibrate 

the  sound  level  to  the  correct  intensity.  Calibration  is  done  with  the  transducers  that  belong  to  the 
audiometer. If a transducer is replaced the audiometer has to be calibrated with the new transducer. 

 
The device has a built-in table for the ISO hearing level offsets (called RETSPL values). These offsets 

are added to the displayed sound level when the device is in normal operation mode, to obtain the ISO 
hearing level. When in calibration mode these offsets are not effective. Therefore all frequencies should 

be  calibrated  to  equal  levels,  namely  80dB.  This  makes  it  easy  and  quick  to  perform  the  calibration. 
Remember to add the correction factors for the calibration microphone. 

 
 

 
Entering calibration mode: 

To  enter  the  calibration  mode,  hold  down  the  ‘ATTENUATOR  DOWN’,  ‘TONE’  and 
‘FREQUENCY  DOWN’  keys  and  switch  on  the  audiometer.  Wait  about  2  seconds  and  release  the 

three buttons. The ‘RESPONSE’ light will now flash five times and then stay on. The audiometer is 
now in calibration mode. 

 