- 13 -

Performing a glucose control solution test

GluNEO

®

 M3 control solution contains a known amount of glucose that reacts with GluNEO

®

 M3 test strip. 

By comparing your control solution test results with the expected range printed on the test strip vial, you 

are able to check that the meter and test strips are working properly, and that you are performing the test 

correctly. It is very important that you perform routine checking to ensure accurate results.

The Glucose control solution should be used

- Before using a new box of test strips.

- Whenever you suspect the blood glucose test meter or blood glucose test strip is not functioning properly.

- If your blood glucose test results are not consistent with your symptoms, or if you think results are not 

accurate.

- If you have dropped the blood glucose test meter.

- For quality control in the point of care usage.

- For teaching or learning the system.

- When using the meter for blood glucose tests after any disinfection procedure.

1. Firmly insert the test strip into the meter test port. Insert deeply in 

direction of arrow on test strip.

Please do not insert the glucose test strip upside down.

2. When you insert the glucose test strip into the glucose test meter, the 

power automatically adjusts the meter accordingly.

3. Press and release either the 

 or 

 

button, then, Control Solution bottle 

icon will appear. Then, press (   ).

제조번호 :

제조일자 :

사용기한 :

품목명

 : 

개인용체외진단검사시약Ⅱ

M3

125

Содержание GluNEO M3

Страница 1: ...Blood Glucose Monitoring System...

Страница 2: ...reprofessional beforemakinganychangestoyour diabetesmanagement TheGluNEO M3BloodGlucoseMonitoringSystemisforin vitrodiagnosticuse only CustomerServiceisavailable24hoursaday 7daysaweek and365daysayear...

Страница 3: ...uNEO M3BloodGlucoseMonitoringSystem DonotchangeyourmedicationbasedontheGluNEO M3metertestresultswithoutcontactingyourphysicianor healthcareprofessionalfirst Donotassembleordisassemblethebloodglucosete...

Страница 4: ...ejectorfunction 22 Reviewingyourresults Reviewingyourbloodglucose 23 Reviewingyouraverages 23 Deletingtestresults 24 Alternatesitetest Lancingandsamplingfromanalternatesitearea 25 27 Lancingdeviceclea...

Страница 5: ...terferewithGluNEO M3teststrips Donotuseduringorsoonafterxyloseabsorptiontesting Xyloseinthebloodwillcauseinterference Beforeusinganyproduct pleasereadallinstructions Consult with your physician or dia...

Страница 6: ...cingDevice optional 4 Lancets optional 5 Instructionsforuse 6 WarrantyRegistrationCard 7 CarryingCase 8 PatientLogbook 9 2EA3VCoinBattery CR2032 YourGluNEO M3BloodGlucoseMonitoringSystemhasbeensealedt...

Страница 7: ...transmitters or other electrical equipment that are sources of electromagnetic radiation as these mayinterferewiththeproperoperationoftheglucosetestmeter DisplayScreen Displaysyourtestresults symbols...

Страница 8: ...splay LowBatteryWarning BeforeHavingaMeal AfterHavingaMeal Pasting BloodDropSymbolforTest MemoryMode SettingMode BluetoothFunction AverageTestResult Warning MeasurementUnit Date Time TestStripCode Ala...

Страница 9: ...besingleuseonly DONOTRE USE Donottestattemperaturesbelow10 C 50 F orabove40 C 104 F Donottestintheconditionthathumidityisbelow10 orabove90 Donotbend cut oraltertheteststrip Avoidgettingdirt food andwa...

Страница 10: ...buttonstoturn theUserActivityOptionon off push button Year Afterpressingand releasingeither or buttons tosettheyear push button Date Time Afterpressingand releasingeither or buttonstosetthedate andtim...

Страница 11: ...ttherighttestunit YourGluNEO M3Meterispre settomg dLastestunit Awrongtestunit will lead to wrong test result display Please consult your healthcare professional if you re not sure which test unit tous...

Страница 12: ...ing thecodenumberontheglucosetestmetereverytime Checkthecodeonthevialtoseeifthecodesmatcheachother CAUTION Donotbendtheglucoseteststriptopreventtheautomaticcoderecognitionfailure If the code recogniti...

Страница 13: ...odglucosetestmeterorbloodglucoseteststripisnotfunctioningproperly Ifyourbloodglucosetestresultsarenotconsistentwithyoursymptoms orifyouthinkresultsarenot accurate Ifyouhavedroppedthebloodglucosetestme...

Страница 14: ...ithinthatrange CAUTION Controlsolutionsisforin vitrodiagnosticuseonly Thismeansthatitisonlyusedfortestingoutsidethebody ControlsolutionisexclusivelymadefortheGluNEO M3BloodGlucoseMonitoringSystem Dono...

Страница 15: ...lsolutiontest Donotuseifitexpired Onceopened thecontrolsolutionexpiresafter6months Ifyourcontrolsolutiontestfallsoutofrange pleasefollowthenextstepsbefore contactingcustomersupport Checktheexpirationd...

Страница 16: ...lood Beingamoreadvancedlancingdevice itprovides ultimatesafetyandcomfortforobtainingbloodsamples Anadjustabletipoffers5differentstrengthlevels ofskinpenetrationforindividualuser scomfort whilealancete...

Страница 17: ...Storage Productsmustbestoredatroomtemperatureprotectedfromsunlightormoisture Cleaninganddisinfection Cleanandwipetheouterpartoflancingdeviceonceaweekattheminimumwithasoftclothdampened withmildsoapandw...

Страница 18: ...hemeteraccordingly 3 Ablinkingblood dropsymbol willappearatthetopofthescreen which indicatesthatthemeterisreadyfortesting CAUTION If the code is not displayed after for a while pull the glucose test s...

Страница 19: ...cets singleuseonly andthelancingdevice Washyourhandsthoroughlywithsoapandcleanwaterafterhandlingthemeter lancingdeviceand orteststrip 3 3 5 1 2 3 4 5 1 3 1 Unscrewthelancingdevicetip 3 3 5 1 2 3 4 5 1...

Страница 20: ...teststripanddonottakeoffyourfingeruntilyougetthebeepsound Thetestwillautomaticallystart Pleaserespectrecommendedquantityofbloodsample Toomuchbloodmaycausemalfunctionofyour meterbycontaminatingit CAUTI...

Страница 21: ...duringsetup pressand release or buttontoselecttheactivity thatcorrelateswith yourresult thenpressandreleasethepowerbutton IfyoudonotselecttouseUserActivityoptionduringsetup pressthe powerbutton Record...

Страница 22: ...teinyourarea Be suretocontactyourlocalrepresentativeforproperdisposal Disposingofyourlancet 1 Pushtheneedleintotheprotectivecover 2 Pullthelancetoutanddiscardaccordingly CAUTION Ifyoupushtheejectorbut...

Страница 23: ...etestmetertocalculate Reviewingyourbloodglucose Pressandrelease button Themostrecentresultappearsfirst Notethetimeanddayofyourblood glucoseresult Pressandreleasethe buttonandyourpreviousresultwillappe...

Страница 24: ...rtblinking Afterthethirdbeepsound youmayhearthatthetestresulthasbeendeleted Deletingallthetestresults Todeletealltestresultsinthememorypress and buttonfor3seconds Thendisplayed willstartblinking After...

Страница 25: ...ralpalmordorsalhand orfingertips IMPORTANT It is recommended that alternative site testing be used when glucoses are stable before meals and before bedtime However whenglucosesarechanging bloodfromthe...

Страница 26: ...nthepalm belowyourthumborpinkyfinger Selecta spotwithoutanyvisibleveinsandawayfromanydeeplines Thismaycause yourbloodsampletosmear IMPORTANT Werecommendthatyoutestonyourfingertipsifyouaretestingforhyp...

Страница 27: ...eplacewiththecleartip 3 Tobringfreshbloodtothesurfaceofthetestsite rubthe testsitevigorouslyforafewsecondsuntilyoufeelitgetting warm Applyingheatmaybehelpful 3 3 5 1 2 3 4 5 1 3 4 Adjustcleartiptotheh...

Страница 28: ...amespotmaycausesorenessand calluses If bruising occurs at an alternate site or you have difficulty getting a sample consider sampling from a fingertip instead Youmaywanttoreviewthechoiceofsiteswithyou...

Страница 29: ...useholdbleachandwaterin1 bleach 9 water ratio Dampenaclothwiththissolutionandwipethelancingdevicethoroughly Soakonlythetipforatleast30 minutesinthedisinfectantsolution Donotsoakthelancingdeviceinliqui...

Страница 30: ...ddryplace Keepawayfromdirectsunlightandheat Donotfreeze Afterremovingateststripfromthevial immediatelycloseittightly Recordthediscardeddate datevialopened 6months ontheteststriplabel Donotuseifitisexp...

Страница 31: ...fter the expiration date printed on the package or vial since it may cause inaccurate results Donottestattemperaturesbelow10 C 50 F orabove40 C 104 F Donottestwithhumiditybelow10 orabove90 Donotbend c...

Страница 32: ...yourmobiledevice selectthemeterthatyouareusing Step4 Ensureyourmeterisswitchedoff thenpresstheleftarrowbuttonpointingupwardsabovetheBluetooth symbolonyourmeterfor3secondsuntilyouhearthesoundofabeep Af...

Страница 33: ...rmobiledevice openyourappandfollowtheinstructionsabove Ifyoudonotwishtotransmityourresultstoyourappatthistime simplypressthemiddle OFF button yourmeterwillentersleepmodeandyourresultscanbetransmittedt...

Страница 34: ...alentlithiumbatteryshouldbeused Follow upaction 1 Checkyourmeterwiththeglucosecontrolsolution 2 Performyourtestagain 3 Ifyouareexperiencingsymptomsthatarenotconsistentwithyourbloodglucosetestresults c...

Страница 35: ...batteriesusingyourlocalbatterycollectionsystemsandincompliance withyourlocalenvironmentallawsandregulations Batteriescontainchemicalsthat ifreleased mayaffecttheenvironmentandhumanhealth Thecrossed ou...

Страница 36: ...testmeter Placethebatteryagain andsetthe bloodglucosetestmeter Iftheproblem persists pleasecontactyourauthorized representativeorcustomersupportat OSANGHealthcareCo Ltd Causedbyeitherusedorwetblood gl...

Страница 37: ...ntactyourlocalrepresentativeorcustomer supportatOSANGHealthcareCo Ltd ifthe problempersists Theambienttemperatureislessthan 10 C 50 F Placethemeterata temperaturebetween10 40 C 50 104 F formorethan30m...

Страница 38: ...ethebatteryandiftheproblem persists contactyourlocalrepresentative orcustomersupportatOSANG HealthcareCo Ltd Theglucosetest meterisnotpowered ON Batteryisdead orthereisaproblemwith themeter Changetheb...

Страница 39: ...on Caution Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment NOTE This equipment has been tested and foun...

Страница 40: ...ngesintopreviouslymanufacturedinstruments 3 OSANGHealthcareCo Ltd hasnoknowledgeoftheperformanceoftheinstrumentwhentheteststrip isalteredormodifiedinanymanner Forwarrantyservice Purchasermustcontactth...

Страница 41: ...t 15 65 Altitude 3048meterUpto 10 000feet OperatingTemperature 10 40 C 50 104 F indoor use only OperatingHumidity 10 90 RH StorageTemperature Store2 30 C 36 86 F andnodirectsunlight DonotFreeze Displa...

Страница 42: ...c medicaldevice Caution consultaccompanyingdocuments Invitrodiagnosticmedicaldevice BatchCode Donotre use Serialnumber Dateofmanufacture Manufacturer Keepawayfromsunlight Wasteelectricalandelectronice...

Страница 43: ...43 MEMO...

Страница 44: ...gu Any ang si Gyeonggi do Korea 14040 www osanghc com OSANG Healthcare Co Ltd Obelis S A Bd Gener al Wahis 53 1030 Brussels Belgiu m Tel 32 2 732 59 54 F ax 32 2 732 60 0 3 E mail mail obelis ne t IN...

Отзывы: