background image

OSAKA

 - OF 33 AZ - OPERATING INSTRUCTIONS - Vr. 01 - PAG. 6

 

If one wishes to set a defrosting cycle every time the instrument 
is switched on (as long as the conditions set in the parameters 
“tSdF” and "tEFE" apply) programme the parameter  

“SdEF”

 = 

yES. 
This  allows  the  evaporator  to  be  permanently  defrosted,  even 
when    frequent  interruptions  to  power  supply  occur  that  may 
cause the cancellation of the various defrosting cycles. 
During  the  defrosting,  it  may  occur  that  the  temperature 
measured  by  the  cell  probe  (Pr1)  increases    excessively  (this 
obviously  depends  on  the  position  of  the  probe  Pr1  compared 
to the evaporator). 
In  the  event  that  one  does  not  wish  this  increase  to  be 
visualised  by  the instrument, it is possible to use the functions 
in  the  parameter

 

“dLo” 

(Block  display  during  defrosting)

 

and

 

“Etdu” 

(Differential unblocking of display after defrosting). 

The  parameter  “dLo”  =  On  allows  the  temperature  Pr1 
visualisation  to  be  blocked  on  the  last  reading  during  a  whole 
defrosting  cycle  and  until  the  temperature  does  not  return 
under the value [”SP” + “Etdu”] after defrosting has finished (or 
the  time  set  in  the  parameter  "dALd"  contained  in  the  block 

]

AL”).  

With  “dLo”  =  Lb,  this  allows  visualisation  of  the  writing 

dEF

 

during defrosting and at the end of defrosting the writing

 PdEF

 

up  to  when  the  temperature  Pr1  does  not  return  below  the 
value [”SP” + “Etdu”] (or the time set on the parameter "dALd" 
contained in the block “

]

AL”). 

Alternatively  with  the  “dLo”  =  OFF,  during  defrosting,  the 
instrument will continue to visualise the temperature measured 
by the probe Pr1. 
Note  that  during  defrosting,  the  temperature  alarms  are 
disabled  during  the  whole  cycle  and  also  afterwards  for  the 
time set in the parameter 

“dALd”

 (see par. 4.9). 

 

4.7 - MANUAL DEFROST

 

To  start  up  a  manual  defrosting  cycle,  press  the  key 
UP/DEFROST when it is not in programming mode and keep it 
pressed  for  about  5  seconds  after  which,  if  the  conditions  are 
correct,  the  led  DEF  will  light  up  and  the  instrument  will  carry 
out a defrosting cycle. 
The  start  up  or  switch  off  commands  of  a defrosting cycle can 
also be given by the digital input that are correctly programmed 
(see par. 4.10). 
 

4.8 - EVAPORATOR FANS CONTROL

 

All  the  parameters  concerning  fan  control  are  contained  in  the 
group 

]

FAn”.

 

The  control  of  the  fans  on  the  output  configured  as  “FAn” 
depending  on  determined  control  statuses  of  the  instrument 
and the temperature measured by the probe Pr2. 
In the case that the probe Pr2 is not used (parameter “Pr 2” = 
OFF)  or  in  error  (E2  o  -E2)  ,  the  output  FAN  is  activated  only 
depending on the parameters “FCOF” and “FEdF”. 
The parameter 

“FCOF”

 decides whether the fans must always 

be  switched  on  independently  of  the  compressor  status 
(“FCOF”=On)  or  be  switched  off  together  with  the  compressor 
(“FCOF”=OFF). 
The  parameter 

“FEdF

” instead decides whether the fans must 

always  be  switched  on  independently  of  the  defrosting  status 
(“FEdF”=On) or switched off during defrosting (“FEdF”=OFF). 
In this latter case, it is possible to delay the start up of the fans 
even  after  the  end  of  the  defrosting  of  the  time  set  in  the 
parameter 

“Fd”. 

When the probe Pr2 is used (par. “Pr 2” = on) the fans, as well 
as  bei

ng  conditioned  by  the  parameters  “FCOF”  and  “FEdF”, 

are also conditioned by the temperature control. 
It  is  possible  to  set  the  disablement  of  the  fans  when  the 
temperature measured by the probe Pr2 is higher than the one 
set  in  the  parameter 

“FLt”

  (temperature  too  hot)  or  when  it  is 

lower than the one set in the parameter 

“Fct”

 (temperature too 

cold). 
The  relative  differential  that  can  be  set  in  parameter 

“dF” 

is 

also associated with these parameters

.

 

d F

d F

t i m e

F A n

o f f

O N

F c t

P r 2

T e m p .

F L t

o f f

o f f

O N

 

 

4.9 - ALARM FUNCTIONS

 

All  the  parameters  concerning  the  alarm  functions  are 
contained in group 

]

AL”.

 

The  alarm  functions  of  the  instrument  work  on  the  internal 
buzzer, if present and configured via the parameter “buF”, and 
on  the  output  desired,  if  configured  by  the  parameters  “O1F”, 
“O2F”  or  “O3F”,    depending  on  what  is  set  on  the  said 
parameters. 
The  possible  selections  of  these  parameters  for  the  alarm 
signalling function are: 

= ALt

 - when one wants  the buzzer or output to be activated in 

alarm  and  can  be  disabled  (alarm  silencing)  manually  by 
pressing  any  key  of  the  instrument  (typical  application  for 
sound signal). 

= AL

 - when one wants  the buzzer or output to be activated in 

alarm status but cannot be disabled manually and are therefore 
only disabled when the alarm status ceases (typical application 
for a light signal). 

= ALL

 - when one wants  the buzzer or output to be activated in 

alarm  status  and  that  they  remain  activated  even  when  the 
alarm  has  ceased  (see  par.4.9.4)  Disablement  (recognition  of 
memorised alarm) can only be carried out manually by pressing 
any key when the alarm has ended (typical application for light 
signal). 

= -ALt

 - when one wants the function described as ALt but with 

an  inverse  function  (buzzer  or  output  activated  in  normal 
condition and disabled in alarm status). 

= -AL

 - when one wants  the function described as AL but with 

inverse  logic  (buzzer  or  output  activated  in  normal  conditions 
and disabled in alarm status). 

=  -ALL

  -  when  one  wants  the  function  described  as  ALL  but 

with inverse working logic (buzzer or output activated in normal 
conditions and disabled in alarm status). 
When no alarms are present, the green LED OK is lit. 
Any  active  alarm  is  shown  on  the  instrument  display  with  the 
lighting up of the AL led and the switching off of the led OK. 
Any  silenced  or  memorised  alarm  status  is  shown  by  the    AL 
led flashing . 
The alarm conditions of the instrument are: 

Probe errors “E1”, “-E1”, “E2, “-E2” 

temperature alarms “HI” and “LO”  

External alarms “AL” 

Open door alarm “AP” 

 

4.9.1 - TEMPERATURE ALARMS

 

The  temperature  alarms  work  according  to  the  probe  Pr1 
measurements,  the  type  of  alarm  set  in  the  parameter 

“Aty”

 

the  alarm  thresholds  set  in  parameters 

“HAL”

  (maximum 

alarm) and 

“LAL”

 (minimum alarm) and the relative differential 

“dAL”.

 

Through  the  parameter  “Aty”  it  is  possible  to  set  the  alarm 
thresholds  “HAL”  and  “LAL”  which  must  be  considered  as 
absolute  (“Aty”=Ab)  or  relative  to  the  active    Set  Point 
(“Aty”=dE). 
Using  some    parameters    it  is  also  possible  to  delay  the 
enablement and the intervention of these alarms. 
These parameters are: 

“PAL”

  -  is  the  temperature  alarm  exclusion  time  on  switching 

on the instrument if the instrument is in alarm status when it is 
switched on. 

Содержание OF 33 AZ

Страница 1: ...5 PROGRAMMABLE PARAMETERS TABLE 6 PROBLEMS MAINTENANCE AND GUARANTEE 6 1 SIGNALLING 6 2 CLEANING 6 3 GUARANTEE AND REPAIRS 7 TECHNICAL DATA 7 1 ELECTRICAL DATA 7 2 MECHANICAL DATA 7 3 MECHANICAL DIMEN...

Страница 2: ...e desired parameter can be selected and pressing the Set key the display will alternately show the parameter code and its setting that can be changed with the UP and DOWN keys Once the desired value h...

Страница 3: ...lectronic devices or electromechanical safety and warning system They should be placed outside the measuring and control equipments possible protective devices responding to specific safety requiremen...

Страница 4: ...y function through a contact that is normally open and closed when the alarm sounds ALt to control a silenceable alarm device through a contact that is normally closed and then open when the alarm sou...

Страница 5: ...abled in WITH HOT GAS or INVERSION OF CYCLE during defrosting the outputs Out and dEF are enabled The automatic defrosting can take place at intervals or if the instrument has an internal clock at set...

Страница 6: ...d When the probe Pr2 is used par Pr 2 on the fans as well as being conditioned by the parameters FCOF and FEdF are also conditioned by the temperature control It is possible to set the disablement of...

Страница 7: ...zer with an alarm memory ALL or ALL it is necessary to set the parameter tAL yES 4 10 DIGITAL INPUT All the parameters concerning the digital input functions are contained in the group din The digital...

Страница 8: ...the second is disabled In this mode the AUX output can be turned off automatically after a certain time that can be set on the parameter tuA With tuA OFF the output is activated and deactivated only...

Страница 9: ...frost 2 OFF 00 00 23 59 hrs min OFF 23 dF 3 Time start defrost 3 OFF 00 00 23 59 hrs min OFF 24 dF 4 Time start defrost 4 OFF 00 00 23 59 hrs min OFF 25 dF 5 Time start defrost 5 OFF 00 00 23 59 hrs m...

Страница 10: ...dEF FAn AuS ALt AL ALL ALt AL ALL Out 61 Out2 Configuration of output function OUT2 see Out1 OFF Out dEF FAn AuS ALt AL ALL ALt AL ALL dEF 62 Out3 Configuration of output function OUT3 see Out1 OFF Ou...

Страница 11: ...250 VAC and OUT3 SPST NO 5A AC1 2A AC3 250 VAC 16 A Max for common pin 1 Electrical life for relay outputs 100000 operat VDE om Installation category II Measurement category I Protection class against...

Отзывы: