PRECAUTIONS (IN USE)
CAUTION
(1) Use the device in enough luminance to read the meter indication.
(2) When the ambient temperature is low, it may happen that the symptoms like
the indications appear on the display panel rather slow, etc. Use the
device in appropriate environmental temperature, 10~40
℃
.
(3) Use an AA-size alkaline dry battery or AA-size nickel-hydrogen battery as the
electric power source. Do not use a manganese dry battery.
(4) Dry batteries are not set up in the device at the time of delivery. Please load
dry batteries properly in the device with the specified manner.
(5) Be careful about electrical leakages due to cervical caries and/or metallic
fillings in the patient's tooth.
(6) Use the K-file or reamer with the plastic handle. The metallic handle can not
be used.
(7) Measure the root canal which is filled with normal saline or sodium hypo-
chlorite as conductive solutions.
(8) Do not use strong acidic water, as there is danger of corroding the device.
●
PRECATIONS (AFTER USE)
(1) After use, wipe both the inside and outside surfaces of the metallic clipping end
of the file holder cord with a cotton dipped in ethanol
●
PRECATIONS (DISINFECTION/STERILIZATION)
(1)Take care of getting sufficient aeration to prevent the side effects from the
remaining gas of EOG sterilization, such as headache, dizziness, nausea, faint,
respiratory arrest, etc. by inhaling. Be careful of the skin wound by touching the
solution.
(2)High pressure steam sterilization is possible for the file holder cord and lip clip
only and if others are given the method of sterilization other than EOG
sterilization, (for example, high pressure steam sterilization), it may cause the
damage to them or they will not be able to perform well with sufficient efficiency.
CAUTION
CAUTION