6
Visual and Audio Indicators
The LCD and audible beeps are designed to provide helpful information
to the user. The screens, symbols, and beeps are explained below.
Compliance Screen
170/185 = 91.9%
Treatment Screen
1:59
Treatment Screen – displays the treatment time remaining
in hours and minutes. The timer counts down to zero until
daily treatment is complete.
Exception Screen
E123
Exception Codes – Display of ERROR, any E codes
(e.g., E01, E02), along with three slow beeps. Contact
Patient Services at 800-535-4492 or 214-937-2718.
Low Battery Warning Screen
1:59
Low Battery – Displays along with three fast beeps when
recharging is recommended.
Charging Complete
Charging Complete – Indicates when the battery is
fully charged.
Battery must be charged
to turn on
E12345678
E12345678
Battery Empty – Indicates that the battery must be
charged before treatment may continue.
Device Expired
Device Expired – Display of a closed lock indicates the
device has been available for treatment for 365 days
and will no longer provide a treatment..
Compliance Screen – Displays a compliance percentage
which is calculated by the number of full treatments days
completed over the number of available treatment days.
The treatments days available begin once the device has
been delivered to the patient and a minute of treatment
time has been established.
Charging Screen
Battery Charging – Battery symbol filling repeatedly
verifies that the device is charging.
Treatment Complete
Daily Prescribed Treatment complete
1-800-535-4492