- Abduction:
41°
- Adduction:
39°
- Internal rotation:
1°
- External rotation:
-1°
- Pauses:
0
- Timer:
continuous operation
- Speed:
100 %
- Reverse on load - motor A:
25
- Reverse on load - motor B:
25
- Motor A:
ON
- Motor B:
ON
- Synchronized mode:
ON
- Total therapy time:
0
- Special functions:
disabled
■
Reverse on load - motor A
(safety circuit)
The motors automatically switch to
opposite directions of movement
when the patient's resistance (load)
exceeds the set value.
Adjustable levels for reverse on load
feature: 1 - 25. At level 1, very low
resistance will cause the device to
reverse; at level 25, a high resistance
is required to initiate the reversal.
Default: level 25
■
Reverse on load - motor B
(safety circuit)
The motors automatically switch to
opposite directions of movement
when the patient's resistance (load)
exceeds the set value.
Adjustable levels for reverse on load
feature: 1 - 25. At level 1, very low
resistance will cause the device to
reverse; at level 25, a high resistance
is required to initiate the reversal.
Default: level 25
■
Motor A ON/OFF
To permit fully isolated movements,
the motors can individually be turned
on and off. Motor A effects adduction
and abduction, motor B effects
internal rotation and external rotation.
For an isolated internal / external
rotation movement, set motor A to
the programmed position (abduc-
tion/adduction), then turn the motor
off.
In normal mode the display indicates
the symbol "
OFF
" during therapy,
instead of the programmed adduc-
tion/abduction angles.
Default: motor A ON
■
Motor B ON/OFF
To permit fully isolated movements,
the motors can individually be turned
on and off. Motor A effects adduction
and abduction, motor B effects
internal rotation and external rotation.
For an isolated adduction/abduction
movement, set motor B to the
programmed position (internal /
external rotation), then turn the motor
off.
In normal mode the display indicates
the symbol "OFF" during therapy,
instead of the programmed inter-
nal/external rotation angles.
Default: motor B ON
Note!
• Please note that one motor (A or B)
must always be turned on. Other-
wise the following symbol will be
displayed when you press
START
:
Caution!
Patient hazard
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked
joints and similar situations. The
manufacturer cannot be held liable for
misuse of this feature.
LEVEL 3:
63
E
n
g
li
s
h