Therefore, avoid direct contact of tissue with the light emission area. 
 

 

Caution: 

Use the automatic control of the light source, if available, or manually adjust the light source to the -
minimum light output required for the application to minimise burn risks.  
 

 

Caution:  

The high light energy emitted at the distal end can lead to high heating or even ignition when the laryn-
goscope is placed on heat-sensitive or combustible surfaces. Therefore, always store the laryngo-
scope in a suitable holder. 
 
Remove dirt from the optical surfaces of the laryngoscope to prevent accidental heat build-up. 
 
Reduce the intensity at the light source or switch it off when the laryngoscope is not in use.  
 
Switch off the light source if the laryngoscope will not be used for a long time! 

 

 

Caution: 

Do not look directly into the light output of the light source or a connected light guide! The energy of 
powerful light sources can cause serious and permanent eye damage. 
 

 

Caution: 

The Video Rhino Laryngoscope may only be used by persons who, on the basis of their training or 
knowledge and practical experience, can guarantee proper handling. 
 

  The operational safety and usability of the medical device depend not only on your skill, but also on 

the care of the device. Regular cleaning and care are therefore necessary (see chapter Cleaning, 
care and disinfection). 
 

  Qualified service and the use of original spare parts give you the guarantee that the operational 

safety, the operational capability and the value of your medical device are maintained. 
 

2  Intended use  

  The Video Rhino Laryngoscope RX1, RS1-PAL, RS1-NTSC is used for endoscopic examina-

tions in human ear, nose and throat medicine. The flexible Video Rhino Laryngoscopes can be 
used to examine the nose, pharynx and larynx, nasal cavities and nasopharynx. The examina-
tion regions can be displayed on high-resolution screens via a (panel) PC. 

 

  The Video Rhino Laryngoscope is not intended for examination of the paranasal sinuses and 

lower larynx. 

 

  It is used exclusively in medical practices and clinics by staff specially trained in handling en-

doscopes.  

 

 

Caution: 

The Video Rhino Laryngoscope RX1, RS1-PAL, RS1-NTSC may only be used on persons who have a 
sufficiently large body opening for the insertion of the insertion tube. This is especially important for 
nasal examinations in children. 

Содержание RS1-NTSC

Страница 1: ...1 Instructions for use Video Rhino Laryngoscope RX1 RS1 PAL RS1 NTSC...

Страница 2: ...ecommended accessories 9 7 1 Cold light source 9 7 2 Monitor 9 8 Connection of the Video Rhino Laryngoscope 9 9 Use of the Video Rhino Laryngoscope 10 9 1 White balance 10 9 2 Insert angle and retract...

Страница 3: ...r and in proper condition including the power supply The Video Rhino Laryngoscope is a high quality precision mechanical optical instrument treat it with care Caution Do not use the Video Rhino Laryng...

Страница 4: ...ge and practical experience can guarantee proper handling The operational safety and usability of the medical device depend not only on your skill but also on the care of the device Regular cleaning a...

Страница 5: ...tructions for use Symbol for separate collection of electrical and electronic equipment Applied part type BF In the instructions for use Caution general danger area IP 68 Unit is dust tight and protec...

Страница 6: ...ge to verify colour fidelity We recommend the use of a lubricant on the shaft before inserting the shaft into the nasal cavity to be examined 4 3 Application The flexible Video Rhino Laryngoscopes fro...

Страница 7: ...ter RX1 RS1 Focus area 10 55 mm Angle of view 90 Diameter distal end 3 9 mm Diameter insertion tube 3 9 mm Distal angulation up down 130 130 Working length 310 mm Total length 540 mm Video format NTSC...

Страница 8: ...nded for use in hospitals without direct connection to the public supply network The RX1 laryngoscope complies with EN 60601 1 2 2007 IEC 60601 1 2 2007 modified with regard to radiation and immunity...

Страница 9: ...o Rhino Laryngoscope 1 Handpiece 2 White balance 3 Capture Button 4 Adjustment lever 5 Light guide plug 6 LED display 7 Supply hose 8 Valve for pressure test 9 Multipoint connector 10 Multipoint socke...

Страница 10: ...hino Laryngoscope 9 1 White balance Attention Before each examination a white balance should be carried out so that the camera repro duces the natural colours Handpiece 1 Handpiece 2 White balance 3 C...

Страница 11: ...disposable cloth afterwards Alternatively we recommend Triamin disinfectant wipes from Wet Wipe available from Orlvision Contact time depending on possible organisms 5 15 minutes see separate data sh...

Страница 12: ...ries from the disinfectant solution with fresh disposable gloves Place the disinfected laryngoscope in a basin tub with microbiologically safe water drinking water quality Use fresh water for each ins...

Страница 13: ...e Disinfection Disinfectant 1 0 Dr Weigert neodisher endo SEPT PAC Automatic cleaning and disinfection process with programme no 22 NORMAL PAA with the following settings Step Pre cleaning for 4 minut...

Страница 14: ...d to the laryngoscope and cannot be pulled off Close the valve on the leak tester Create a test pressure of 160 10 mmHg by pumping the leak tester Figure 7 Connection pressure tester Caution If the ma...

Страница 15: ...provides the authorised companies with all necessary product documentation Caution Unauthorised opening repairs and modifications to the laryngoscope release orlvision from any liabil ity for operatio...

Страница 16: ...These instructions for use have been machine translated Orlvision GmbH Gewerbestrasse 17 D 35633 Lahnau Tel 49 0 6441679298 0 Fax 49 0 6441679298 99 info orlvision com www orlvision com Rev 25 25 11 2...

Отзывы: