76
BT37 IFU
5.
Specifications
5.1
System specifications
Dimensions
420 mm wide x 270 mm deep x 210 mm high.
Weight
15.5 kg
Storage temperature
-10 °C to +50 °C
Storage humidity
5% to 95% relative humidity non-condensing
Storage special instructions
Recharge every 4 months by connecting to the mains power supply
for 24 hours.
Operating environment
For indoor use only
Operating temperature
+5 °C to +40 °C for safe operation
See the table Control on page 77 for control limitations..
Operating humidity
5% to 90% relative humidity non-condensing
Altitude
Up to 2000 m
Pollution degree
Pollution degree 2 (BS EN61010-1)
IP rating
IP31
Electromagnetic environment
The equipment is intended for use in a basic electromagnetic
environment, characterized by being supplied directly at low voltage
from the public mains network.
Содержание BT37
Страница 1: ...MA101414 ML BT37 IFU 0088 EN...
Страница 2: ......
Страница 3: ...3 I...
Страница 15: ...15 Left arrow decrements selected digit Enter key moves to the right...
Страница 21: ...21 II...
Страница 24: ...24 BT37 IFU Blow through pipes with medical grade gas to clear any foreign bodies...
Страница 25: ...25 Do not over tighten...
Страница 26: ...26 BT37 IFU...
Страница 27: ...27 Do not over tighten Connect up to 10 incubators in series Check for leaks...
Страница 29: ...29 Remove cap Refit cap Ensure tubes aligned correctly Fit filter...
Страница 35: ...35 2 3 External data collection 2 4 Alarm connection...
Страница 36: ...36 BT37 IFU 2 5 Connecting to the mains supply...
Страница 37: ...37 III...
Страница 40: ...40 BT37 IFU 3 2 Setting the password Set passwords to prevent unauthorised access...
Страница 43: ...43 3 3 1 1 1 Changing the left hand chamber temperature...
Страница 47: ...47 Flow increased for purge duration after lids are shut...
Страница 55: ...55 3 4 Changing the humidifier...
Страница 56: ...56 BT37 IFU See Installing the humidifier on page 28...
Страница 57: ...57...
Страница 58: ...58 BT37 IFU 3 5 Switching off 3 6 Battery backup...
Страница 59: ...59...
Страница 60: ...60 BT37 IFU...
Страница 61: ...61 IV...
Страница 65: ...65 4 2 2 Disinfecting Clean first see page 63 Disconnect mains power Spray surfaces with 70 IPA Wait 15 min...
Страница 67: ...67 4 3 Safety testing Check for leaks...
Страница 69: ...69 4 6 Troubleshooting...
Страница 70: ...70 BT37 IFU...
Страница 71: ...71...
Страница 73: ...73 4 9 Disposal Do not dispose of with general waste Clean and disinfect first see page 63...
Страница 74: ...74 BT37 IFU...
Страница 75: ...75 V...