Operation
Babysense2 is ready for operation and use once the installation is
complete. To operate, follow the instructions below:
• Place your baby in cot.
• Turn control unit on by pressing On/Off button.
• Red control lights switch on briefly and a double click sound is heard.
• When breathing and movement is detected, the blue light flashes.
• Acoustic and visual red alarm light are activated if baby's breathing
movement rate stops or slows significantly (below 10 breaths per
minute).
• If red lowbattery warning light flashes, replace all batteries as soon as
possible. If an additional clicking sound is heard, all batteries must be
replaced immediately.
• Turn off by pressing the On/Off button before taking your baby out of
cot.
• Failure of the sensor pads to be directly against the mattress. A
mattress of a stiff material such as polystyrene foam could cause false
alarms.
Testing Operation
You must conduct this quick performance test daily.
• The performance test should be conducted while baby is not in cot.
• Turn device on and take two steps back from cot or leave the room.
• Ensure that alarm goes off within 20 seconds (due to absence of
breathing movements).
• The alarm may sound with or without preliminary ticking.
• Check that alarm can be heard in all locations throughout the home or
place of care.
• In the case that the alarm does not sound within 20 seconds refer to
troubleshooting on page 14.
• This performance test must be conducted every day.
Operation