The holder of the UV tubes is accessible now.
Turn the retaining ring and remove it from the tube. You can easily access the
tube now.
Detach the tube carefully from the holder. A narrow tool (e.g. screwdriver)
can be helpful to loosen the tube from the holder.
The tube can be removed from the unit now.
Pull the retaining ring over the new tube to the end of the plug.
Insert the new tube with retaining ring into the unit.
Be sure, that the lamp tube end will be positioned correctly into the matching
holder. The left service cover can also be removed to have a better view of
the holder.