Recirculating Air Disinfection System V-lab
orca
GmbH
www.
uvpro
.de
Seite 10 von 28
3.
ELECTRONIC COMPONENTS
Electronic components
The
UV
pro
V-LAB Recirculating Air Disinfection System has a high-quality
ballast (10)
for optimum
operation of the
UV
pro
tube. In this way maximum power, high frequent ignition and high durability
can be guaranteed. The high-frequency technology ensures a very high efficiency and low power
consumption. Of course, all components are in conformity with the “Low Voltage Directive” and “EMC
directive”.
If a tube has a malfunction or in case of a defect in the wiring to or in the tube socket the ballast
automatically switches the connected tube off. Thus, the circuit board will be protected and a
malfunction can be detected by our monitoring units. The temperature of the square tube with the
fully encapsulated ballast is a maximum of 35 °C (at room temperature). The optimum temperature is
between 0 °C and +40 °C. Heating to 70 °C in a short term is possible but leads to a momentary power
loss and permanently accelerates the aging of the system.
Connection directly or with UV
pro
monitoring unit
- Connection with the grid
230 V
AC
±10 %
. The protection of the connection via a
residual current
circuit breaker
(30 mA) is mandatory.
- UV
pro
CU1
The optional monitoring unit
CU1
can be used for one
UV
pro
device. You
will find further information in the documentations of this monitoring
unit.
- UV
pro
UVC-Line 8/16/24
The optional monitoring unit
UVC-Line 8/16/24
can be used for several
UV
pro
devices. Further information can be found in the documentations
of this monitoring unit.
Содержание UVpro V-LAB
Страница 1: ...UVC Recirculating Air Disinfection System V LAB Operating Instruction ...
Страница 2: ......
Страница 13: ...Recirculating Air Disinfection System V lab orca GmbH www uvpro de Seite 11 von 28 Dimensions and sketch ...
Страница 28: ...Recirculating Air Disinfection System V lab orca GmbH www uvpro de Seite 26 von 28 ...
Страница 29: ...Recirculating Air Disinfection System V lab orca GmbH www uvpro de Seite 27 von 28 ...