Orantech EtCO2 Скачать руководство пользователя страница 8

8

9

•  Interfering Gas and Vapour Effects

Gas or Vapor

GAS Level

Quantitative Effects

Nitrous Oxide

60

No additional effect

Halothane

4

No additional effect

Enflurane

5

No additional effect

Isoflurane

5

No additional effect

Sevoflurane

5

No additional effect

Xenon

80

CO2 value at 38 mmHg, the 
maximum negative bias 
Poor 5mmHg

Helium

50

No additional effect

Metered Dose Inhaler Propellants

unspecified unspecified

Desflurane

15

CO2 value at 38 mmHg, the 
maximum positive bias Poor 
3mmHg

Ethanol

0.1%

No additional effect

Isopropanol

0.1%

No additional effect

Acetone

0.1%

No additional effect

Methane

1%

No additional effect

Содержание EtCO2

Страница 1: ...User Manual EtCO2 Sensors The difference you can sense...

Страница 2: ...que for monitoring CO2 The principle is based on the fact that CO2 molecules absorb infrared light energy at specific wavelengths with the amount of energy absorbed being directly related to the CO2 c...

Страница 3: ...ch sensor is individually packaged Sensors must be stored in their original packaging and within the storage conditions to maximize the storage life Storage conditions are as follows Ambient temperatu...

Страница 4: ...ad The airway adapter clicks into place when seated correctly See figure 1 Set the sensor to zero Install the airway adapter at the proximal end of the circuit between the elbow and the ventilator Y s...

Страница 5: ...l the airway adapter at the proximal end of the circuit between the elbow and the ventilator Y section see figure 3 For intubated patients with an integrated airway adapter in the breathing circuit Co...

Страница 6: ...atient as shown and then attach the oxygen supply tubing to the oxygen delivery system and set the prescribed oxygen flow see figure 6 Alarm Limits Ranges and Default Settings Refer to the User manual...

Страница 7: ...2 mmHg 41 70 mmHg 5 of reading 71 100 mmHg 8 of reading 101 150 mmHg 10 of reading Above 80 breaths per minute 12 of reading NOTE Gas temperature at 25 C Respiration Rate Range 0 to 150 breaths per m...

Страница 8: ...onal effect Sevoflurane 5 No additional effect Xenon 80 CO2 value at 38 mmHg the maximum negative bias Poor 5mmHg Helium 50 No additional effect Metered Dose Inhaler Propellants unspecified unspecifie...

Страница 9: ...a dry cloth and leave to dry completely Caution Do not sterilize by irradiation steam or ethylene oxide Warranty Liability Orantech offers 12 months warranty against defects in material or workmanshi...

Страница 10: ...Incompatible components can result in degraded accuracy and performance Consult the operation instructions for the equipment concerned and the related accessories before operating equipment to ensure...

Страница 11: ...ice to sale by or on the order of a physician Waste Disposal Please refer to your local laws and regulations for information on how to dispose of EtCO2 sensors Title of Symbol Manufacturer Caution Dat...

Страница 12: ...ral rubber latex Non sterile Refer to instruction manual booklet Type BF Applied Part Gas Port Input Federal U S Law restricts this device to sale by or on the order of a physician Gas Port Output REF...

Отзывы: