Technical Data
11
-
2.4 Air Purge Laminar
The lens must be always kept clean from dust, smoke, fumes, and other contaminants in order to avoid reading
errors. These effects can be reduced by using an air purge laminar. Make sure to use oil-free, technically clean
air only.
The needed amount of air (approx. 20...100 l/min.) depends on the application and the installation
conditions on-site. Pipe connection of air purge laminar is 6x4 mm.
2.5 Protective window
Protective window
Description
Spectral range
Transmission
1)
Germanium
Protective window (50,8 x 3 mm/ Ge) for PI 450i G7, 640i G7
7,9 µm
0.96
Protective window (50,8 x 3 mm/ Ge) for PI 400i, 450i, 640i
8-14 µm
0.92
Zinc sulfide
Protective window (50,8 x 3 mm/ ZnS) for PI 450i G7, 640i G7
7,9 µm
0.93
Protective window (50,8 x 3 mm/ ZnS) for PI 400i, 450i, 640i
8-14 µm
0.91
Protective window (50,8 x 3 mm/ ZnS) for LT models of CTlaser-, CSlaser serie
8-14 µm
0,91
Borofloat
Protective window (50,8 x 3 mm/ Borofloat 33) for PI 1M, 08M, 05M
0,85-1,1 µm, 780-820 nm,
500-540 nm
0.92
1)
The displayed values are standard values and can vary between different delivery units.
When using the foil
[part number: ACOPHSF]
, the transmission must be set on 0.85 (deviations possible).
Содержание Outdoor protective housing
Страница 8: ... 8 2 2 Dimensions Figure 1 Dimensions mm with wall mount ...
Страница 9: ...Technical Data 9 Figure 2 Dimensions for wall mount ...
Страница 10: ... 10 2 3 Electrical Connections Figure 3 Connection circuit board Power supply Industrial PIF and USB server ...
Страница 15: ...Installation 15 The USB server is now ready mounted Figure 9 Figure 9 USB server ready mounted ...
Страница 17: ...Installation 17 Step 2 Carefully remove the mounting plate Figure 11 Figure 11 Remove the mounting plate ...
Страница 26: ... 26 Figure 24 Remove air purge laminar ...
Страница 32: ... 32 Appendix A Declaration of Conformity ...
Страница 33: ......
Страница 34: ...optris Outdoor protective housing MA E2022 10 A ...