Digital Command Set
Read commands
Header bytes
Response
Conversion Response to Decimal value
Example
read process temperature
1)
3E0200
word (hiByteLobyte)
process temp [°C] = (Hex
Dec(word)-1000)/10
[1]
read head temperature
3E0202
word (hiByteLobyte)
head temp [°C] = (Hex
Dec(word)-1000)/10
read current target temperature
1)
3E0204
word (hiByteLobyte)
current temp [°C] = (Hex
Dec(word)-1000)/10
read current ambient temperature
3E0206
word (hiByteLobyte)
ambient temp [°C] = (Hex
Dec(word)-1000)/10
read current emissivity
3E0208
word (hiByteLobyte)
emissivity = Hex
Dec(word)/1000
[2]
Set commands
Header bytes
Set value
Generation of the set value
set emissivity
3A0208
word (hiByteLobyte)
word = Dec
Hex (emissivity x 1000)
[3]
switch on loop maintenance mode
3D026190
------
------
[4]
set target temperature for maintenance
3A0212
word (hiByteLobyte)
word = Dec
Hex (target temperature [°C] x 10 +1000)
[5]
switch off loop maintenance mode
3D026180
------
------
[6]
Examples
Send
Receive
Comment
[1] read process temperature
3E0200
0519
process temp [°C] = (Hex
Dec(
0519
)-1000)/10 = 30,5
[2] read current emissivity
3E0208
036C
emissivity = (Hex
Dec(
036C
)/1000) = 0,876
[3] set emissivity to
0,95
3A0208
03B6
------
word = Dec
Hex(
0,95
x 1000) =
03B6
[4] switch on loop maintenance mode
3D026190
------
------
[5] set analog output to
0
°C (permanent)
3A0212
03E8
------
word = Dec
Hex (
0
[°C] x 10 +1000) =
03E8
[5] set analog output to
200
°C (permanent)
3A0212
0BB8
------
word = Dec
Hex (
200
[°C] x 10 +1000) =
0BB8
[6] return to standard mode
3D026180
------
------
Burst string
Example
Complete burst string
Conversion to Decimal value
2 synchronisation bytes: AAAA
------
------
2 bytes for each output value (hi lo)
03B8
AAAA 03B8
process temp [°C] = (Hex
Dec(
03B8
)-1000)/10 =
-4,8
After switch on a continuous serial signal will be created. The burst string can be configured with CompactConnect software.
Communication mode (bidirectional)
Burstmode (unidirectional)
1)
if
peak/ valley hold is activated
the "process temperature" holds the detected peak or valley whereas the "current target temperature" shows the real process
temperature (without post processing); in
standard mode
"process temperature" and "current ta
optris CS – E2010-12-A
26