12 Trigger knob

 

13 Current 

14 Life time

 

15 Reset

 

16 Lamp box connector

 

17 Power interface

 

18 Fuse cap

 

19 Power switch of mercury lamp power source

 

 

LED Fluorescence Device Operational Parts Marking Graphs 

 

 
 

 

 
 

 

 
 

 

 
 

 

 
 

 

 

20 Lamp condenser adjusting lever                      21 LED switch shift lever 
22 brightness adjusting knob                                  23 power switch 
24 charging indicator                                                  25 fuse tube   
( Single group LED lamp source is similar as two group lamp source)   

 
1)

 

Insert the plug of the mercury power source into external power supply socket 
(please make sure the rated current and voltage are coincided with the input 
supply voltage at first).

 

2)

 

Turn off the Epi-fluorescence switch and turn on the mercury power source 
(the input voltage should be within 100V

220V). Mercury Power Box should 

be opened by pressing start button (START or TRIGGER)to excitation mercury 
lamp,it takes 10 minutes to make the mercury lamp reach the stable state and 
max. luminescence efficiency. 

3)

 

Put  10X  fluorescence  objective  in  the  optical  path  and  lower  transmission 
condenser to a minimum or take off. 

4)

 

Place the fluorescence specimen on the stage and fix it with the clamp, adjust 
the stage knobs to move the specimen in the optical path. 

5)

 

Pull the filter converting lever to the needed position. 

16

17

 

18

 

19

 

14

13

12

15

 

21 

20

22

23

24

25

Содержание A16.2601-L

Страница 1: ...A16 2601 L LED Fluorescent Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument ...

Страница 2: ...ions 2 1 Applications 3 2 Principle 3 3 Instruction 4 4 Installation 5 5 Fluorescence Device Operation 8 6 Use of fluorescent centering device 11 7 Fluorescent mercury lamp replacement 12 8 Replacement of the fuse 12 ...

Страница 3: ...lammable materials can t near the lamp in case of fire 5 Notes On Replacing The Bulb Replacement should be based on the identity of the instrument using the same specifications of the bulb otherwise it may cause equipment damage The power supply must be cut off before bulb replacement the bulb must be cooled off completely before proceeding 6 Carry Power must be cut off before moving Be careful no...

Страница 4: ...ple and standard provided with four group excitation filters system blue B green G violet V and ultraviolet UV The light emitted from the lamp is converted to the excitation light e g blue light with specified wavelength by going through the excitation filter then passes through dichroic prism and objectives the objective plays role of condenser to irradiate vertically the object The object is exc...

Страница 5: ... mercury lamp light source field adjustable diaphragm and aperture adjustable diaphragm switch fluorescence module by turret device standard equipped with B G Uv and V FL2 two groups fluorescence device also equipped with mercury lamp light source field adjustable diaphragm and constant aperture adjustable diaphragm aperture adjustable diaphragm is optional turret conversion two group module and b...

Страница 6: ...e locking screw 4 fluorescence device main body 5 observing tube mount 6 light box locking screw 7 mercury lamp light box 8 connector wire for mercury lamp box 9 protective plate screw 10 connector for main body 11 wrench 12 power box output socket 13 external power connector 14 connector socket for mercury lamp box 9 10 11 12 13 Mercury Power Box 5 2 3 4 6 7 8 1 14 5 4 3 2 1 6 7 8 14 ...

Страница 7: ...r connector 1 Take all parts from the package remove the protective package and place it on the vacant bench 2 Turn fluorescence device and put the protective plate in with the screw then tighten it by wrench 3 Then put the device in upright direction then link the main body connector with main body bayonet fasten the microscope with screw 4 Slightly loosen the locking screw then connect the tube ...

Страница 8: ...nd relevant socket 7 According to the installation instruction to install microscope main body Rechargeable Lithium power source specially for LED illumination Eyepiece tube connect with fluorescence device Fluorescence device connect with microscope stand Connect with microscope stand Tube connector Main body Observing tube ...

Страница 9: ...vice Installation Marking Graphs 1 Fluorescence module switch driver 2 Field diaphragm centering handle 3 Field diaphragm adjusting lever 4 Aperture diaphragm centering handle 5 Aperture diaphragm adjusting lever 6 Lamp condenser adjusting lever 7 Screw for opening mercury lamp 8 Light source vertical adjusting knob 9 Light source horizontal adjusting knob 10 Light switch lever 1 2 3 5 6 4 7 8 9 1...

Страница 10: ...voltage are coincided with the input supply voltage at first 2 Turn off the Epi fluorescence switch and turn on the mercury power source the input voltage should be within 100V 220V Mercury Power Box should be opened by pressing start button START or TRIGGER to excitation mercury lamp it takes 10 minutes to make the mercury lamp reach the stable state and max luminescence efficiency 3 Put 10X fluo...

Страница 11: ...0 and the indicator is flashing as the the battery capacity is 80 The indicator is bright as the battery capacity is full The indicator comes dark after the external power supply cut Before perform fluorescence observation locate the specimen with the transmission light first To prevent the fluorescence from attenuation quickly block the excitation light with barrier when preparing for fluorescenc...

Страница 12: ... the fluorescence device to G and watch the light source s position through the widow c Adjust the handle of light source condenser properly and make the outline of the light source clear adjusting the knob on fluorescence lamp house to center d The cross line center in observation window must be aligned with the lamp house Please turn the fluorescence device to G because the brightness of G is su...

Страница 13: ...forbidden to take the lamp holder out of the lamp house while the mercury lamp lights on The mercury lamp should not be replaced until it cools down completely Clean the bulb with the gauze soaked with a few mixture of alcohol and ether 4 6 Stains such as dust and fingerprint are not allowed on the bulb surface 8 Replacement of the fuse Cut off the power supply screw off the fuse cap on the power ...