8
-
12 Trigger knob
13 Current
14 Life time
15 Reset
16 Lamp box connector
17 Power interface
18 Fuse cap
19 Power switch of mercury lamp power source
LED Fluorescence Device Operational Parts Marking Graphs
20 Lamp condenser adjusting lever 21 LED switch shift lever
22 brightness adjusting knob 23 power switch
24 charging indicator 25 fuse tube
( Single group LED lamp source is similar as two group lamp source)
1)
Insert the plug of the mercury power source into external power supply socket
(please make sure the rated current and voltage are coincided with the input
supply voltage at first).
2)
Turn off the Epi-fluorescence switch and turn on the mercury power source
(the input voltage should be within 100V
~
220V). Mercury Power Box should
be opened by pressing start button (START or TRIGGER)to excitation mercury
lamp,it takes 10 minutes to make the mercury lamp reach the stable state and
max. luminescence efficiency.
3)
Put 10X fluorescence objective in the optical path and lower transmission
condenser to a minimum or take off.
4)
Place the fluorescence specimen on the stage and fix it with the clamp, adjust
the stage knobs to move the specimen in the optical path.
5)
Pull the filter converting lever to the needed position.
16
17
18
19
14
13
12
15
21
20
22
23
24
25