background image

                                                                                       

A15.0908 Series

 

(2) Push the light source group base

 into the 

reflected illuminator light source base

 thoroughly, 

then lock the lock screw

 

2-2-3

 Assemble the Condenser (With Polarizer Group) 

For Transmitted Illumination 

(1) Rotate the coarse focusing knob

  to raise stage 

to the highest. (See Fig. 6) 
(2) Rotate the condenser up-down knob

 to move 

the condenser bracket to the lowest. 
(3) Loosen the condenser lock-screw

 fully. 

(4) Swing out the front lens of condenser

 with the 

scale forward. Make the lock screw

 of condenser 

in alignment with the groove

  of the condenser base. 

Push the condenser into the innermost of base. 
(5) Screw down the condenser lock screw, and raise 
the condenser to the highest position with the 
condenser up-down knob.

 

 

                                             

                                     

 

 

2-2-4

 Assemble the Stage Clip

 

Plug the two stage clip pins

 into the hole

 on 

rotatable stage thoroughly. (See Fig.7) 
When using the moving stage, plug the two 

φ

3mm 

forelocks into the two holes

, and lock them on 

stage with M4 screws. 

★ 

Adjust the handle to “60” reticle (to “0”) when 

using the moving stage on X-axis, to avoid 

hurting objectives. 

 

 

2-2-5

 Assemble the Objective

 

Rotate the coarse focusing knob to lower the stage. 
Screw the 10X objective into the hole of nosepiece 
with red mark, then install the objectives into the 
nosepiece from the lowest magnification to the 
highest in a clockwise direction. (See Fig.8)   
 

Fig.6

Fig.7

 

Fig.5 

Fig.8

 

Содержание A15 0908 Series

Страница 1: ...and maintenance to the A15 0908 Series Polarizing Microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modificat...

Страница 2: ...e cap 10 3 7 Adjust the Objective Center 10 3 8 Center the Condenser 11 3 9 Adjust the Field Diaphragm 12 3 10 Adjust the Aperture Diaphragm 12 3 11 Adjust the Analyzer 13 3 12 Adjust the Polarizer 13...

Страница 3: ...ed only 12V 50W Halogen bulb 7 Wide voltage range is supported as 100 240V Additional transformer is not necessary Make sure the voltage is in this range 8 Use the special wire supplied by our company...

Страница 4: ...er Adjustment Ring Epi Illuminator Group Upper Light Source Group Middle Head Bertrand Lens Centering Screw Objective Nosepiece Light Adjusting Knob Objective Body Coarse Adjusting Limit Knob Fine Foc...

Страница 5: ...ss Flashboard Group Polarizer Flashboard Group Stage Lock Screw Tension Adjustment Ring Stage Clip TV and Photography Device Lock Screw Light Path Selecting Pole Trinocular Head Condenser Center Adjus...

Страница 6: ...es 2 1 Assembling Scheme Following is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust or dirt Assemble carefully and do not scrap any par...

Страница 7: ...ert the pins into the receptacles Make sure the bulb is vertical See Fig 3 3 Loosen the bulb lock sheet to lock the bulb and put on the cover group Before replacing cut off the main power supply and w...

Страница 8: ...ser into the innermost of base 5 Screw down the condenser lock screw and raise the condenser to the highest position with the condenser up down knob 2 2 4 Assemble the Stage Clip Plug the two stage cl...

Страница 9: ...ddle head into the hole on the Epi Illuminator with a little left inclined and then screw down the lock screw The screw head should be pushed into the V slot on the right side of the middle head coatt...

Страница 10: ...on plug into the aviation socket of the microscope 4 Insert one end of the power cord into the microscope power socket 5 Insert the other end of the power cord into the power supply socket Cancel Step...

Страница 11: ...t the switch to and push the spectroscope pole to the innermost to move the spectroscope into the light path then adjust the light adjustment knob until the illumination is comfortable for observation...

Страница 12: ...cusing knob for clear details 3 3 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane after focusing or the stage declines itself please...

Страница 13: ...avoid damaging to both glasses and eyepieces 2 Open the eye cap if the user doesn t wear glasses so that it can prevent stray light disturbing the observation See Fig 20 3 7 Adjust the Objective Cente...

Страница 14: ...g 23 Move the front lens of the condenser into light path when use the objective beyond 20X 3 Rotate the 20X objective into the light path and focus the specimen 4 Rotate the field diaphragm adjustmen...

Страница 15: ...d of view then the field diaphragm is centered correctly See Fig 26 5 In actual use enlarge the field diaphragm a little to make its image circumscribed to the field of view For Transmitted Illuminati...

Страница 16: ...the analyzer is shifted into the light path See Fig 29 2 Loosen the lock screw turn the analyzer turntable to change the vibrated direction of the analyzer A grid value of scale is 2 and the adjustin...

Страница 17: ...an Bertrand lens and make the vibration direction of polarizer be vertical with the analyzer 3 14 Use the Compensator The compensator is inserted into the slot of the middle head in 45 declined The su...

Страница 18: ...a then turn the coarse and fine focusing knobs until the image is clear 3 Rotate the Bertrand lens switch turntable locate it on position B to move the Bertrand lens into light path At that time the b...

Страница 19: ...nermost for binocular observation pull the light path selecting pole to the outermost for trinocular observation TV and photography See Fig 35 3 19 2 Assemble and Use the TV Device 1 Loosen the lock s...

Страница 20: ...pole to the outmost then operate the photography device according to its instructions 3 20 Use the Digital Viewing Head 3 20 1 Connect the Power Cord Insert the output connector of the transformer in...

Страница 21: ...sabled when the white balance switch is popup while it is enabled when the switch is pressed down The switch is mainly for adjusting the color of the microscope TV image See Fig 40 When using the digi...

Страница 22: ...k or not even Stain or dust has accumulated on the lens objectives or eyepieces Clean the lens Stains have accumulated on the specimen Clean the specimen 4 Stain or dust is observed in the view field...

Страница 23: ...slide is not fixed correctly Adjust it correctly 5 Cannot move the slide smoothly The movable specimen holder is not fixed properly Fix it correctly 3 Electrical System No power supply Check the conn...

Страница 24: ......

Отзывы: