background image

24

 

 

Chapter 6 Troubleshooting               

 

                              A16.1065

 

6.1

 

Optical System

 

TROUBLE

 

CAUSE

 

SOLUTION

 

The edge of the field 

of  view  is  dark  or 

the  brightness  is  not 

uniform

 

The nosepiece is not in the located 

position (objective and light path not 

coaxial)

 

Locate the nosepiece properly 

where it clicks

 

 

 

A lens (the objective, condenser, 

eyepiece or collector) is dirty.

 

Clean it thoroughly

 

Dirt 

or 

dust 

is 

visible in the field of 

view

 

A  lens  (the  objective,  condenser, 

eyepiece or collector) is dirty.

 

Clean it

 

Dirt/dust on the specimen

 

Clean it

 

   

Specimen is placed    reversely 

Turn it over

 

A  lens  (the  objective,  condenser, 

eyepiece or collector) is dirty. 

Clean it

 

The aperture diaphragm is too large 

Appropriate down

 

Dirt or dust on the eyepiece

 

Clean it

 

The aperture diaphragm is too small

 

Appropriate up

 

One side of image is 

blurred

 

Condenser is not properly centered

 

Center the condenser with the 

centering screw

 

The nosepiece is not properly 

engaged

 

Engage the nosepiece properly

 

The specimen is not clamped

 

Clamp it with stage clips

 

The  image  moves 

while  focusing  on 

the zoom

 

Specimens float on the surface of the 

platform 

placed it stably 

The nosepiece is not properly 

engaged

 

Engage the nosepiece properly

 

The brightness is not 

enough

 

The aperture diaphragm is too small

 

Appropriate up

 

 

Содержание A14.1065

Страница 1: ...B G U Instruction Manual This instruction manual is intended for users of the Inverted Series Biological Microscope A16 1065 To ensure safety use perform the best performance and to familiarize yours...

Страница 2: ...Specimen Placement and Stage Knob Adjustment 10 3 3 1 Mechanical Stage 10 3 3 2 Pallet 11 3 3 3 Auxiliary Stage 11 3 4 Adjusting the Interpupillary Distance 11 3 5 Focusing on the Specimens 12 3 5 1...

Страница 3: ...5 Installing the Eyepiece 20 4 2 6 Installing the Camera 20 4 3 Assembly Episcopic Illumination Microscopy 21 4 3 1 Installing the Filter Cube 21 4 3 2 Installing the Fluorescence LED Unit 21 4 3 3 R...

Страница 4: ...event lens glass to stick to fingerprint perspiration other factors influence observation prevent lens and other attachment drop damaged 2 Avoid placing the microscope in place with direct sunlight hi...

Страница 5: ...e rear of the main body Do not hold any other parts such as the upper part of the illumination pillar focus knobs eyepiece tube or stage when carrying the microscope Doing so might result in dropping...

Страница 6: ...use Cover up must wait for the lamp room to cool down 9 The microscope usage environmental requirements a For indoor use b The environment temperature range 10 35 c Maximum relative humidity 80 at 31...

Страница 7: ...tive 4 Phase contrast objective 10 Phase contrast objective 20 Phase contrast objective 40 Infinity plane semi apochromatic objective 10 Infinity plane semi apochromatic objective 20 Infinity plane se...

Страница 8: ...Magnification 4 10 20 40 10 20 40 numerical aperture N A 0 10 0 25 0 40 0 60 0 30 0 45 0 60 Work distance mm 30 10 2 12 2 2 7 4 8 3 6 Cover glass thickness 0 17 0 17 0 17 0 17 1 2 1 2 1 2 Conjugate d...

Страница 9: ...A16 1065 1 2 Components The microscope A16 1065 components following diagrams A16 1065 Viewing head Camera mount LCD Mainframe Nosepiece Stage Condenser Objective Contrast slider Moving stage Eyepiec...

Страница 10: ...nents following diagrams A16 1065 FL Viewing head Camera mount LCD Mainframe Fluorescence illumination Stage Condenser Filter Cube turret Fluorescent baffle Eyepiece Nosepiece Objective Moving stage U...

Страница 11: ...adjustment knob 7 Condenser brightness 2 Condenser rotating knob 8 Fine focus knob 3 Condenser adjustment knob 9 Coarse focus knob 4 aperture diaphragm adjusting knob 10 Power indicator 5 Nosepiece 1...

Страница 12: ...9 A16 1065 1 Bracket horizontal movement knob 2 Bracket vertical movement knob 3 Fluorescence cube turret A16 1063 3 1 2...

Страница 13: ...careful when changing the objective The objective may collide with the specimen because of the short working distance of specimen observation 3 3 1 Movving stage The Movving stage is mounted and used...

Страница 14: ...er 3 3 3 Auxiliary stage Connect the auxiliary stage to the left and right of the stage to expand the stage width After connecting the auxiliary stage the stage width is 300 mm 3 4Adjusting the Interp...

Страница 15: ...or about 4 mm trip to limit The traveling distance of the objective for each knob is as follows Rotation of knob Distance traveled by objective Fine focus knob 1 rotation 0 2 mm Coarse focus knob 1 ro...

Страница 16: ...result in an appropriate contrast and a favorable image The iris aperture adjusting handle is marked with the aperture of the diaphragm 3 7 Nosepiece Turn the nosepiece until it clicks in place stays...

Страница 17: ...aphragm image with the phase ring image in the objective Decentering the phase ring and the PH annular diaphragm Basically the phase ring and the PH annular diaphragm should be adjusted so that they a...

Страница 18: ...and switch the filter cube by rotating the filter cube switching unit 3 9 3 Filter cube switching unit Rotating the filter cube switching unit switches the filter cube When the filter cube is switched...

Страница 19: ...on filter BP filter a barrier filter BA filter and a dichroic mirror DM Referring to the following items as a guide select a desired combination of filter cubes according to the characteristics of the...

Страница 20: ...minute scale figures flash then click the button hour scale figures become flashing When the time figures flash three times it means setting successful Double click locking brightness or unlocking Whe...

Страница 21: ...h component is shown in the following figure and the digital indicate the installation step Save the supplied hexagonal spanner You need it when you change parts CAUTION Before installation make sure...

Страница 22: ...contrast slider horizontally into the condenser socket and notice the adjusting screw side of the phase contrast slider is facing the operator side Push into the process when you hear the click sound...

Страница 23: ...itching the objective rotate the objective until it clicks to ensure that the desired objective enters the center of the optical path 4 2 5 Installing the Eyepieces Insert the eyepiece into the eyepie...

Страница 24: ...scope and remove the cover 2 Mount the LED unit corresponding to the desired wavelength at the position that matches the address of the filter cube The back cover of the LED unit is marked with positi...

Страница 25: ...ff immersion oil from objectives Do not use petroleum benzene to wipe the prism surface of the eyepiece tube or filters 5 1 1 Cleaning the Lenses Keep the lenses free of dust fingerprints and other di...

Страница 26: ...disposal If a sample comes into contact with this product check the sample is hazardous or not If the sample is hazardous follow the standard procedure of your laboratory 5 2 Storage Store this produ...

Страница 27: ...ced reversely Turn it over A lens the objective condenser eyepiece or collector is dirty Clean it The aperture diaphragm is too large Appropriate down Dirt or dust on the eyepiece Clean it The apertur...

Страница 28: ...llary Distance does not adjust Adjust the Interpupillary Distance correctly the illumination is not appropriate Adjust illumination switch 6 3 Electrical System TROUBLE CAUSE SOLUTION The bulb cannot...

Отзывы: